Advisor
®
Tech Vital Signs Monitor
6
Inputting Patient ID Information
1.
Once in the Patient ID area, set the
ID
number of the patient (should be
unique
)
2.
Enter patient
name
(up to 9 characters)
3.
Enter patient
type
(canine, feline, equine, bovine, ovine, caprine or swine)
4.
Enter patient
sex
5.
Set patient
blood type
(default = N/A, A, B, O, AB, RH+, RH-)
6.
Set patient
age
(default = 25)
7.
To exit, choose
Save
for new patient data to be saved
PATIENT SPECIFIC PARAMETER SET UP
Starting a New Case
1.
Touch
Patient ID
in the top left area of the display screen directly and enter new
information (outlined in the next section)
2.
Or, touch the
Pause
soft-key and then choose
Start New Case
It is important to start a new case
for every patient
to ensure that
unique readings, memories and
trends can be enabled for
accurate
vital signs monitoring.
Setting Unique Parameters for Your Patient
Each individual parameter can be adjusted for your patient needs by touching the
desired parameter on the right side of the display screen.
(When selecting each parameter, a description is displayed at the bottom of the screen.)
ECG
1.
Alarm Switch
: Adjustable (default = off)
2.
HR Alarms
: Select to change high and low values
3.
ECG Lead
: Select appropriate lead (default = I)
4.
ECG Gain
: Increase or decrease size of waveform
5.
ECG Filter
: Select from the options below (default = surgery, which is the most
common filter)
•
Monitor: Used for basic surgeries (not for any surgery which includes electric i.e.
cautery, laser, dental scaling)
•
Surgery: Used in instances with cautery, laser, dental scaling; assists in
filtering HZ
•
Diagnostic
6.
Lead Type
: Default = 5 leads (most common is 3 leads)
7.
ECG Cascade
: It is possible to cascade ECG for longer monitoring times (set up must
be 6 waveforms)
8.
Sweep Speed
: Adjustable (default = 25 mm/s)
9.
Sweep Color
: Adjustable (default = green)
10.
ARR Analysis
: Turn arrythmia analysis on or off
Did you know that you can customize
sweep colors?
Sweep colors can be chosen from
white, gray, red, yellow, green, cyan,
blue and magenta.