Chapter 5: Operating Instructions
5-8
Home Use Operation Manual
Upon power-up, the following parameters will be set to the last values saved before power-down:
SpO
•
2
averaging time
Pulse rate averaging setting
•
Trend interval
•
Device mode
•
Printer output format
•
SpO
•
2
alarm limits (See Note below)
Pulse rate alarm limits
•
Alarm volume (See Note below)
•
Pulse beep on/off volume
•
Normal/High Sensitivity setting
•
nOTe: If the low SpO
2
limit is set to less than 85% it will be reset to 85% when the monitor
is next powered on. The high SpO
2
limit will be adjusted to 86% if it is 85% or less.
nOTe: If the alarm volume is set to a value less then 8 at power down, it will be reset to 8
when the monitor is next powered on.
Checking the Monitor’s Performance
Pulse oximeters do not require user calibration. If checking the function of the device is desired,
an Oximetry Patient Simulator (Smiths Medical PM, Inc. catalog number 1606) is available as an
accessory. The simulator attaches to the oximeter in place of the sensor. It provides a known SpO
2
and pulse rate signal to the oximeter.
nOTe: A Patient Simulator does not calibrate the monitor. The monitor does not require
calibration. A Patient Simulator provides a known SpO
2
value and pulse rate to the
monitor that allows the monitor’s performance to be checked.
nOTe: A Patient Simulator cannot be used to assess the accuracy of a pulse oximeter and/
or sensor.
nOTe:
f
Follow the instructions included with the Patient Simulator.
Содержание Oxi-Pulse 30
Страница 2: ......
Страница 6: ...Table of Contents iv Home Use Operation Manual This page is intentionally left blank ...
Страница 16: ...Chapter 1 Introduction 1 10 Home Use Operation Manual This page is intentionally left blank ...
Страница 22: ...Chapter 3 Getting Started 3 4 Home Use Operation Manual This page is intentionally left blank ...
Страница 46: ...Chapter 6 Alarms 6 6 Home Use Operation Manual This page is intentionally left blank ...
Страница 62: ...Chapter 9 Troubleshooting 9 6 Home Use Operation Manual This page is intentionally left blank ...
Страница 65: ......