background image

PRODUCT(S) DESCRIBED MAY NOT BE LICENSED OR AVAILABLE FOR SALE IN CANADA AND OTHER COUNTRIES

Smiths Medical India Private Limited

508-509, Western Edge 2,

W.E. Highway, Borivali (E),

Mumbai- 400066. India

T: +91 22 28705210 F: +91 22 28705211

E: csindia

@

smiths-medical.com

www.smiths-medical.com
Find your local contact information at: www.smiths-medical.com/customer-support/contact-us 

Smiths Medical is part of the global technology business Smiths Group plc. Please see the Instructions for Use/Operator’s Manual for a complete listing of the indications, contraindications, warnings 

and precautions. HOTLINE, Level 1 and the Smiths Medical design mark are trademarks of Smiths Medical. The symbol ® indicates the trademark is registered in the U.S. Patent and Trademark 

Office and certain other countries. All other names and marks mentioned are the trademarks or service marks of their respective owners. The product referenced is CE-marked.  

©2019 Smiths Medical. All rights reserved. TM194875GBIN-082019

MPAUCA-1371

For more information visit our website at www.smiths-medical.com

2 6

Troubleshooting*

No Power:

• 

Confirm the HOTLINE

®

 warmer is plugged  

in properly

• 

Confirm the power switch in in the ON position (note: if the 
HOTLINE

®

 warmer is plugged in and the power switch is 

turned ON, the green or red LED will illuminate)

Check Disposables Alarm:

• 

Confirm the HOTLINE

®

 fluid warming set is properly installed

• 

Push the Twin-Tube Connector firmly into the socket on 
the right side of the HOTLINE

®

 warmer (note: turn OFF 

the power switch before replacing the HOTLINE

®

 fluid 

warming set)

• 

If the alarm is not cleared, replace the HOTLINE

®

 fluid 

warming set. Turn ON the power switch and verify the 
alarm has cleared

• 

If the alarm is not cleared, remove the HOTLINE

®

 warmer 

from service

Add Recirculating Solution Alarm:

• 

Check the level in the reservoir – turn OFF the power switch, 
remove the HOTLINE

®

 fluid warming set if installed, and add 

recirculating solution to the maximum level

Over Temperature Alarm:

• 

Check the HOTLINE

®

 fluid warming set for kinks or other 

restrictions

• 

Check for air lock:

• 

Turn the power switch to OFF, remove the HOTLINE

®

 fluid 

warming set, and gently shake the HOTLINE

®

 warmer to 

remove air

• 

Plug in the HOTLINE

®

 fluid warming set and turn the 

power switch ON

• 

If the alarm is not cleared, remove the HOTLINE

®

 warmer 

from service and return for repair or replacement

Electrical Interference (receiving or transmitting):

• 

Move the HOTLINE

®

 warmer away from  

the device in question

• 

Plug the HOTLINE

®

 warmer into a separate electrical circuit 

(if the problem continues, notify Smiths Medical or your local 
Smiths Medical distributor)

If the suggested solutions do not correct the problem, 
discontinue use of the HOTLINE

®

 fluid warmer, remove from 

service, and contact Smiths Medical.

HOTLINE

®

 Fluid Warmer (HL-90) 

HOTLINE

®

 Fluid Warming Sets (L-70)

Step-by-Step Guide