PRODUCT(S) DESCRIBED MAY NOT BE LICENSED OR AVAILABLE FOR SALE IN CANADA AND OTHER COUNTRIES
Smiths Medical India Private Limited
508-509, Western Edge 2,
W.E. Highway, Borivali (E),
Mumbai- 400066. India
T: +91 22 28705210 F: +91 22 28705211
E: csindia
@
smiths-medical.com
www.smiths-medical.com
Find your local contact information at: www.smiths-medical.com/customer-support/contact-us
Smiths Medical is part of the global technology business Smiths Group plc. Please see the Instructions for Use/Operator’s Manual for a complete listing of the indications, contraindications, warnings
and precautions. HOTLINE, Level 1 and the Smiths Medical design mark are trademarks of Smiths Medical. The symbol ® indicates the trademark is registered in the U.S. Patent and Trademark
Office and certain other countries. All other names and marks mentioned are the trademarks or service marks of their respective owners. The product referenced is CE-marked.
©2019 Smiths Medical. All rights reserved. TM194875GBIN-082019
MPAUCA-1371
For more information visit our website at www.smiths-medical.com
2 6
Troubleshooting*
No Power:
•
Confirm the HOTLINE
®
warmer is plugged
in properly
•
Confirm the power switch in in the ON position (note: if the
HOTLINE
®
warmer is plugged in and the power switch is
turned ON, the green or red LED will illuminate)
Check Disposables Alarm:
•
Confirm the HOTLINE
®
fluid warming set is properly installed
•
Push the Twin-Tube Connector firmly into the socket on
the right side of the HOTLINE
®
warmer (note: turn OFF
the power switch before replacing the HOTLINE
®
fluid
warming set)
•
If the alarm is not cleared, replace the HOTLINE
®
fluid
warming set. Turn ON the power switch and verify the
alarm has cleared
•
If the alarm is not cleared, remove the HOTLINE
®
warmer
from service
Add Recirculating Solution Alarm:
•
Check the level in the reservoir – turn OFF the power switch,
remove the HOTLINE
®
fluid warming set if installed, and add
recirculating solution to the maximum level
Over Temperature Alarm:
•
Check the HOTLINE
®
fluid warming set for kinks or other
restrictions
•
Check for air lock:
•
Turn the power switch to OFF, remove the HOTLINE
®
fluid
warming set, and gently shake the HOTLINE
®
warmer to
remove air
•
Plug in the HOTLINE
®
fluid warming set and turn the
power switch ON
•
If the alarm is not cleared, remove the HOTLINE
®
warmer
from service and return for repair or replacement
Electrical Interference (receiving or transmitting):
•
Move the HOTLINE
®
warmer away from
the device in question
•
Plug the HOTLINE
®
warmer into a separate electrical circuit
(if the problem continues, notify Smiths Medical or your local
Smiths Medical distributor)
If the suggested solutions do not correct the problem,
discontinue use of the HOTLINE
®
fluid warmer, remove from
service, and contact Smiths Medical.