background image

• 

2

HOTLINE

®

 Fluid Warmer (HL-90) 

HOTLINE

®

 Fluid Warming Sets (L-70)

Step-by-Step Guide

• 

Plug the Twin-Tube Connector on the HOTLINE

®

 fluid 

warming set into the socket

• 

Connect the IV fluid and the intravenous administration set to 
the HOTLINE

®

 fluid warming set

• 

Fully prime the intravenous administration set and the 
HOTLINE

®

 fluid warming set prior to connecting to the patient

• 

After use, turn OFF the power switch

• 

Remove the HOTLINE

®

 fluid warming set and insert the reflux 

plug into the socket

• 

Turn ON the power switch

• 

Inspect the patient end of the HOTLINE

®

 fluid warming set for 

leaks to confirm the integrity of the intravenous pathways

IMPORTANT: Please refer to the Operator’s Manual supplied 

with the product for detailed instructions, contraindications, 
warnings and cautions and the appropriate IFU for specific 
information on using disposable sets.

Warning: Prime the recirculating solution path before 

connecting to the intravenous extension set, to confirm there 
is no breach between the recirculating solution path and the 
intravenous path.

• 

1

• 

3

• 

4

HL-90

L-70

Отзывы: