S E C T I O N 1 3
•
S p e c i f i c a t i o n s a n d A c c e s s o r i e s
40
HOTLINE
®
Blood and Fluid Warmer | Operator’s Manual
Electromagnetic Environmental Recommendations
Accessories
Recommended separation distances between portable and mobile RF communications
equipment and the HOTLINE
®
Warmer
The HOTLINE
®
Warmer is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the HOTLINE
®
Warmer can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the HOTLINE
®
Warmer as recommended below, according
to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
d=[3.5/V1]
P
80 MHz to 800 MHz
d=[3.5/E1]
P
800 MHz to 2.5 GHz
d=[7/E1]
P
0.01
0.116
0.116
0.233
0.1
0.368
0.368
0.737
1
1.16
1.16
2.33
10
3.69
3.69
7.38
100
11.66
11.66
23.33
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P
is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people
.
Product Description
L-70
HOTLINE
Fluid Warming Set with Injection Port
L-70NI
HOTLINE
Fluid Warming Set without Injection Port
L-80
HOTLINE
Warming Set with L-10 Gas Vent
L-10
Gas Vent
PC-8
T-Connector, 20.3 cm (8”) Patient Lead with Injection Port
YC-8
Y-Connector, 20.3 cm (8”) Patient Lead with Injection Port
Содержание LEVEL 1 HOTLINE HL-90-40
Страница 1: ...OPERATOR S MANUAL l s HOTLINE Blood and Fluid Warmer HL 90 40 6 4524104 Rev 000 3 07 40 C Set Point ...
Страница 2: ...Page intentionally blank ...
Страница 3: ...HOTLINE Blood and Fluid Warmer HL 90 40 OPERATOR S MANUAL PN 4534010EN Rev 004 s l ...
Страница 53: ...Page intentionally blank ...
Страница 54: ...s l H Manufacturer Smiths Medical ASD Inc 160 Weymouth Street Rockland MA 02370 USA Tel 1 800 258 5361 ...