Chapter 4: Operating Instructions
Clinician’s Operation Manual
4-7
Turning On the Monitor
Figure 4-6: Turning On the Monitor
To turn on the monitor, press the
x
key. Before using the monitor, check the following at power-
up:
All LEDs light.
•
The monitor’s software revisions (main, battery PIC, oximeter) are momentarily displayed.
•
The patient record number is momentarily displayed.
•
After a few seconds the % SpO
2
value, pulse rate, pulse signal strength and PI bar graphs should
be shown. If not, see
Chapter 9: Troubleshooting
for help.
WARNING! Verify that all LEDs (light emitting diodes) on the display light up upon startup
of the device.
If the oximeter detects an error during power up, “
ERR
” will display in the pulse rate section of
the display. A numeric error code will display in the SpO
2
section. See
Chapter 9: Troubleshooting
for more information.
Press this key to turn
the monitor on and off.