2. Insert the cassette hooks into
the hinge pins on the pump.
3. Place the pump upright on a
firm, flat surface. Press down
so the cassette fits tightly
against the pump.
4. Insert a coin into the latching
button, push in, and turn
counterclockwise until the
mark on the latch lines up
with the solid dot.
WARNING: Attach the cassette (the part of the CADD™
Medication Cassette Reservoir or CADD
®
Administration Set
that attaches to the pump) properly. An improperly attached or
detached cassette could result in unregulated gravity infusion
of medication from the fluid container or a reflux of blood,
which could result in death or serious injury.
5. Gently twist and pull on the cassette
to make sure it is firmly attached.
6. A message will appear on the
display showing the type of
cassette you have latched. Press
„
.
2000-03-08 D. Zurn
«Prism Attach Cass Table»
2000-03-08 D. Zurn
«Prizm Twist Cass»
High Volume Admin
set latched
NEXT to continue