4
2008
VVP-15B
ELECTROFISHER
SET-UP INSTRUCTIONS
1. Set the POWER switch to the OFF or
down position.
2. Set the OUTPUT VOLTAGE ADJUST
control to the extreme counter-clockwise
position (Labeled MIN.).
3. Set the MODE SELECT switch to the
OFF position.
4. Connect the electrodes to be used to the
desired output connector.
5. Connect the AC INPUT 230 VAC
connector to the 240V AC 60 Hertz
power source (generator).
6. Set the METER SELECTOR switch to
the desired position.
7. Set the OUTPUT (range) switch to the
desired position. (AC mode must be in
300V range).
8. Position the MODE SELECT switch to
desired AC or DC mode.
9. Select desired DUTY CYCLE &
FREQUENCY, (AC or Cont. DC are
selected at this step when such output is
desired)
BEGIN ELECTROFISHING
1. Toggle power circuit breaker ON.
2. Depress foot switch or anode pole thumb
switch.
3. Slowly increase output by rotating OUTPUT
ADJUST CONTROL in slow clockwise
movement.
4. During operation, before changing OUTPUT
(range) switch, MODE switch, or selected
electrode output from one output to another,
set the OUTPUT ADJUST control to the
extreme counter-clockwise position (MIN.)
and release footswitch(s)/pole switch(s).
WARNINGS
WARNING!: HIGH VOLTAGE IS USED IN THE OPERATION OF THIS EQUIPMENT!
DEATH ON CONTACT MAY RESULT IF PERSONNEL FAIL TO OBSERVE SAFETY PRECAUTIONS.
BE CAREFUL TO AVOID CONTACT WITH ALL CIRCUIT COMPONENTS AND CONNECTIONS
WHILE CHECKING OR SERVICING THIS EQUIPMENT
This equipment will not function properly when operated from a 240 Volt AC generator if the gen-
erator neutral line is grounded or connected to any other points. Consult generator manufacturer
for specifications regarding whether your generator has a floating neutral or procedures to modify
your generator. This applies to both the 240Volt and 120V outputs. Additional caution should be
taken to insure that any devices used on the 120V output are not internally grounded to the neu-
tral.
If the neutral connections cannot be removed a 240 volt to 240 volt isolation transformer of adequate
wattage rating must be used
(5000-Watt Nominal, SRI line #7791).