12
3. At the End of A Daily Treatment Period:
The device will automatically stop treatment after 20
minutes. A two-tone beeping sound will indicate that
treatment has been completed, and the LCD and system
will then turn OFF automatically.
a. Remove the transducer from the RAF. The transducer
must be removed from the RAF after each treatment
session.
b. Using a soft cloth, tissue, paper towel or a cotton
swab, gently clean the skin area at the treatment
site to remove the coupling gel (
Figure 7
). Clean
the inside surface of the RAF and remove any gel
that may have accumulated on the transducer to
prevent gel hardening on the device.
c. Replace the round felt plug with tab, into the RAF
(
Figure 8
)
.
d. Snap on the RAF cap by positioning it over the RAF
hole and pushing down until it snaps securely in place.
G.Interruption of a Treatment Period
It is recommended that each treatment period be a continuous,
uninterrupted 20-minute period. During a treatment period,
if for any reason you must attend to something for a short
Figure 7
Pulsing Treating Symbol indi-
cates device is operating
properly.
Treatment countdown will
continue as long as proper
treatment is occurring.
Countdown will pause if gel
error occurs.
Figure 6a