- 10 -
4 FUNCTIONING
4.1
FIRST START UP
ATTENTION
The first start up of the system must be carried out by the installer who will make sure
the machine runs properly.
Insert the feed plug in a 230V 50Hz outlet equipped with efficient grounding. Now the
system is able to work. It advisable to run the run for at least a half an hour before
drinking it, in order to allow the system to eliminate the substances used to conserve the
reverse osmosis membrane.
Make sure there are no leaks in the circuit and fasten the antiflooding probe on the
bottom of the machine.
Close the cover with the screws position the machine in its final destination and push the
brakes of the whells.
Operate sanitation of equipment as described in section 5.8.
The WO-03 membranes are provided saturated in a maintenance solution. Before taking water, let it run
for at least 15-30 minutes.
4.2 NORMAL
USE
In order to normally use WO-03 the machine must always be kept on.The machine starts automatically at
every rising cycle of the dish/wash machine. Every six hours the machine automatically opens the inlet
solenoid valve, in this way the membranes are washed for 2 minutes.
To adjust the salinity of the water, acting on the valve for mixing later.
If high water consumption occurs, the by-pass solenoid valve opens and raw water flous to the outlet.
ATTENTION
Absolutely avoid letting the membranes dry out; this could irreparably compromise its
functioning.
4.3 INACTIVITY
ATTENTION
It is advisable, each time the machine is inactive for a period over one week, to let the
water run for at least 10 minutes before using it.
It is advisable to sanitize the system every six months during routine maintenance
procedures.
Contact assistance in order to carry out this operation, which must be performed by
specialized personnel/installation technicians.
4.4
ELECTRONIC BOARD OPERATING
4.4.1 Normal functioning
When the card is in normal functioning mode, the name of the reseller is
displayed.
When the faucet is opened, the machine starts to produce permeate water;
pushing the key ENTER
, the conductivity is displayed.
SMEG
uS/cm
0050