Pressurizing and starting up the unit:
After all pipes have been connected, pressurise
the system gradually and check for leaks.
Turn the “Osmosis pump” switch to “
I
”. Check
that the following message displays:
In normal operation the value
Rj
varies during
osmosis-treated water production.
Wait
5-10 minutes,
then continue by starting up
the delivery pump.
Osmosis pump switch
I M P O R T A N T !!!
Delivery pump startup (first startup)
Remove the red bleeder cap
P
(“Sfiato innesco”)
on the front (close to the pressure gauge,
labelled as such) .
To remove the cap, press the John Guest clip
and allow a bit of water to run into a container;
refit the cap
P
, and actuate the “Delivery pump”
switch.
FAILURE TO FOLLOW THIS PROCEDURE TO
THE LETTER CAN CAUSE SERIOUS
DAMAGE TO THE DELIVERY PUMP.
Delivery pump switch Cap “P”