HRX-OM-X084
Chapter 9 Documents
HRR Series
9.1
Specifications
9-1
Chapter 9 Documents
9.1 Specifications
9.1.1 HRR010-A-10/20-
Table
9-1
Specifications [HRR010-A-10/20-(DMT1)U(YZ)]
Model
HRR010-A-10-(DMT1)U(YZ)
HRR010-A-20-(DMT1)U(YZ)
Cooling method
Air-Cooled refrigerated type
Refrigerant
R134a(HFC)
R410A(HFC)
Quantity of refrigerant
kg
0.24
0.22
Control method
PID
control
Ambient temperature andhumidity and altitude
1
Temperature:
5 to 40
℃
,
Humidity
:30 to 70%,
Altitude
:
Less than 3000m
Circulating
fluid system
Circulating fluid
2
Tap water, Ethylene glycol aqueous solution 15%
Operating temperature
range
1
℃
15 to 35
Cooling capacity
(50/60Hz)
3
W
770 / 950
950 / 1100
Heating capacity
(50/60Hz)
4
W
340 / 430
250 / 300
Temperature stability
5
℃
±0.1
Pump capacity
(50/60Hz)
6
MPa
0.09(at 5L/min) / 0.11(at 5L/min)
For option
–T1 : 0.35(at 5L/min) / 0.35(at 5L/min)
Rated flow rate
(50/60Hz)
7
L/min
5 / 5
For option
–T1
:
5 / 5
Flowdisplay range
20
L/min
2 to 16
Electric conductivity
display range
µS/cm
0.1 to 48(For option -DM)
Electric conductivity
setting range
µS/cm
0.5 to 45(For option -DM)
Particle filter nominal
filtration rating
19
µm
35
Bypass valve
19
Installed
Tank capacity
L
Approx.
3
Fluid outlet,fluid return port port size
Rc1/2
Drain port size
Rc1/4,With cap
Leakage protection
Drain pan (With water leakage detector
19
)
Fluid contact part material
Stainless steel,Copper brazing (Heat exchanger)
12
,
Aluminum oxide ceramic,Carbon,PP,PE,PPE,POM,PET,PA,FKM,EPDM,NBR,
PVC,PPS,fluoropolymer
13
,Ion exchange resin
13
Electrical
system
Power supply
1-phase AC100V (50/60Hz)
1-phase AC115V (60Hz)
Allowable voltage fluctuation±10%
15
1-phase AC200 to 230V (50/60Hz)
Allowable voltage fluctuation±10%
15
Circuit protector
A
15
10
Applicable earth leakage breaker
8
Rated current
:
15A
Sensitivity current
:
30mA
Rated current
:
10A
Sensitivity current
:
30mA
Cable Qty x size(Inculding ground)
16
3 cores ×14AWG(3 cores×2.0m
㎡
)
Rated operating current
(50/60Hz)
17
A
6.2 / 7.7
2.5 / 3.0
For option
–T1
For option
–T1
9.5 / 10.7
4.3 / 4.7
Rated
power consumption
(50/60Hz)
17
kW
(kVA)
0.50 / 0.67
(0.56 / 0.70)
0.48 / 0.60
(0.51 / 0.61)
For option
–T1
For option
–T1
0.72 / 0.84
(0.86 / 0.96)
0.73 / 0.80
(0.85 / 0.94)
Communication function
Contact input / output, Serial RS-485 / RS-232C
Noise level (50/60Hz)
9
dB(A)
60 / 60
59 / 59
Dimensions
10
mm
W483xD550xH221
Accessory
18
Power supply connector, Operation manual, Particle filter element
19
Weight
11
kg
32
29
1: Use the product in conditions where freezing will not occur. Consult with SMC if using in a season or region where
the ambient temperature will fall below zero.
2: If tap water is used, use water which satisfies the standard of The Japan Refrigeration And Air Conditioning
Industry Association (JRA GL-02-1994/Cooling water system
–circulation type - make-up water).
3: (1)Ambient temp: 25
℃
, (2)Circulating fluid temp.: 20
℃
, (3)Circulating fluid rated flow, (4)Criculating
fluid : Tap water,(5) Power supply :100V type:AC100V, 200V type :AC200V, (6)Piping length : Shortest
The cooling capacity will be reduced by 300W when option T1 [Inverter pump] is selected.
4: (1)Ambient temp: 25
℃
, (2)Circulating fluid temp:20
℃
,(3) Circulating fluid rated flow, (4)Criculating fluid : Tap water,
(5) Power supply :100V type:AC100V, 200V type: AC200V, (6)Piping length : Shortest
Содержание HRR010-A-10 Series
Страница 14: ...HRX OM X084 Chapter 1 Safety Instructions 1 8 Material Safety Data Sheet MSDS HRR Series 1 8...
Страница 46: ...HRX OM X084 Chapter 3 Transport and Setting Up 3 5 Fill of circulating fluid HRR Series 3 24...
Страница 120: ...HRX OM X084 Chapter 6 Option 6 5 Option Z HRR Series 6 6...
Страница 141: ...HRX OM X084 Chapter 8 Control Inspection Exchange and Cleaning HRR Series 8 4 Replacement of consumables 8 15...
Страница 146: ...HRX OM X084 Chapter 8 Control Inspection Exchange and Cleaning 8 5 Electrical schematic diagram HRR Series 8 20...
Страница 165: ......