background image

High-Pressure Pump Mounted

Option symbol

Possible to choose a high-pressure pump in accordance with customer’s piping resistance.
Cooling capacity is reduced by approx. 20 W due to the heat generation from the pump.

P

HEF002 A6

P

High-pressure pump mounted

   Options have to be selected when 

ordering the thermo-con.
It is not possible to add them after 
purchasing the unit.

HEF 

Series

Option

Air-cooled

10

Содержание HEF Series

Страница 1: ...94 s a 41 s Compared with the existing product HECR002 Time to change the circulating fluid temperature from 30 C to 20 C when the circulating fluid IN and OUT are directly connected Now only takes ab...

Страница 2: ...se design Noise Level Sound level Noise level reference Sound example 20 dB Whisper The sound of leaves touching each other or the sound of snow falling 30 dB Very quiet Suburbs late at night or the s...

Страница 3: ...emperature sensor and out from the circulating fluid outlet The heat dissipation air flow direction is as shown in Figure 2 Air is taken from the front of the Thermo con and discharged from the rear D...

Страница 4: ...m Bumper Circulation Between Gas Generation Analyzer and JacketedVessel Application Examples Peltier type Chiller Thermo con Compact Type Air cooled HEF Series C mount lens F mount lens Microscope C m...

Страница 5: ...ecifications p 6 Cooling Capacity p 7 Heating Capacity p 7 Pump Capacity Thermo con Outlet p 7 Dimensions p 8 Operation Display Panel p 9 Alarm p 9 Maintenance p 9 Option High Pressure Pump Mounted p...

Страница 6: ...down to 20 C Water bath Example When cooling the object below a certain temperature in certain period of time Cooled substance total volume V 2 L Cooling time h 15 min Cooling temperature difference...

Страница 7: ...4 Fluid contact material Stainless steel EPDM NBR Ceramics PPE PPS Carbon PP POM Electrical system Power supply 24 VDC 10 Power consumption 12 5 A Peak current 18 A Alarm Refer to Alarm Communication...

Страница 8: ...rox 20 W For the high pressure pump Option the heating capacity increases by approx 10 W Pump Capacity Thermo con Outlet Circulating fluid Water 0 18 0 16 0 14 0 12 0 10 0 08 0 06 0 04 0 02 0 0 1 1 5...

Страница 9: ...ignal input 5 SG 6 Output cutoff alarm open when alarm occurs 7 Common for output cutoff alarm 8 RUN STOP signal input 9 RS 485 BUS Power Cable Accessory Connector J42FSC 02V KX J S T Mfg Co Ltd made...

Страница 10: ...be repaired at one of SMC s sites As a rule SMC will not conduct on site maintenance Alarm Alarm code Alarm description Operation status Main reason AL0 Memory error Stop There is an abnormality with...

Страница 11: ...customer s piping resistance Cooling capacity is reduced by approx 20 W due to the heat generation from the pump P HEF002 A6 P High pressure pump mounted Options have to be selected when ordering the...

Страница 12: ...ecifications Power supply input 100 240 VAC Single phase 50 60 Hz Max 3 9 A DC output 24 VDC Max 12 5 A Operating environment Ambient temperature 0 to 40 C Humidity 20 to 80 RH no condensation DC outp...

Страница 13: ...s less than these values the thermo con will not be able to provide precise control and the repeated cooling and heating operations may cause it to fail 7 A magnet driven pump is used as the circulati...

Страница 14: ...Copper Cu mg L 0 1 or less Sulfide ion S2 mg L Should not be detected Ammonium ion NH4 mg L 0 1 or less Residual chlorine Cl mg L 0 3 or less Free carbon CO2 mg L 4 0 or less 11 If there is a point w...

Страница 15: ...of SMC products with production equipment for the manufacture of weapons of mass destruction WMD or any other weapon is strictly prohibited 2 The exports of SMC products or technology from one countr...

Страница 16: ...c hu Ireland 353 0 14039000 www smcautomation ie sales smcautomation ie Italy 39 03990691 www smcitalia it mailbox smcitalia it Latvia 371 67817700 www smc lv info smc lv Lithuania 370 5 2308118 www s...

Отзывы: