After removing the solidified powder, perform ultrasonic cleaning and autoclave sterilization.

2

1

Water Leakage

( Refer to the Operation Manual 7-3~7-4 )

Replace the O-ring(s) 
at the handpiece joint 
section of the powder 
case, apply a drop of 
the supplied turbine oil 
and rub it in with your 
fingers.

Troubleshooting

Powder Clogging

Nozzle

Powder Case

Cleaning wire (small)

Cleaning wire (large)

Cleaning wire (large)

Internal Nozzles

Handpiece

Cleaning file

Cleaning file

• Basic maintenance methods after use are described in the Quick Maintenance 

Manual. Refer to the operation manual for more detailed instruction.

• Keep the Operation Manual and the Quick Maintenance Manual for future 

reference.

• Perform ultrasonic cleaning once a week to prevent clogging of powder. (Refer 

to the Operation Manual 8-1)

• Thoroughly dry the inside of the Powder Case and the connection part after 

each maintenance process.

Caution

( Refer to the Operation Manual 8,10 )

Remove the Nozzle from the Handpiece and clean the inside of the Nozzle and Handpiece with the supplied Cleaning Wires 
and Cleaning File. In the same way, clean the Handpiece Joint Section and the Internal Nozzles of the Powder Case.

To avoid bending, do not apply excessive 
force to the Cleaning wire or file during use. 