After removing the solidified powder, perform ultrasonic cleaning and autoclave sterilization.
2
1
Water Leakage
( Refer to the Operation Manual 7-3~7-4 )
Replace the O-ring(s)
at the handpiece joint
section of the powder
case, apply a drop of
the supplied turbine oil
and rub it in with your
fingers.
Troubleshooting
Powder Clogging
Nozzle
Powder Case
Cleaning wire (small)
Cleaning wire (large)
Cleaning wire (large)
Internal Nozzles
Handpiece
Cleaning file
Cleaning file
• Basic maintenance methods after use are described in the Quick Maintenance
Manual. Refer to the operation manual for more detailed instruction.
• Keep the Operation Manual and the Quick Maintenance Manual for future
reference.
• Perform ultrasonic cleaning once a week to prevent clogging of powder. (Refer
to the Operation Manual 8-1)
• Thoroughly dry the inside of the Powder Case and the connection part after
each maintenance process.
Caution
( Refer to the Operation Manual 8,10 )
Remove the Nozzle from the Handpiece and clean the inside of the Nozzle and Handpiece with the supplied Cleaning Wires
and Cleaning File. In the same way, clean the Handpiece Joint Section and the Internal Nozzles of the Powder Case.
To avoid bending, do not apply excessive
force to the Cleaning wire or file during use.