Bed Pad System
TEST THE SYSTEM
• Plug the pad into the monitor and apply pressure
to pad to activate the system. The monitor will beep twice
when active.
• The alert will sound when pressure is released
from the pad. Press the “reset” on the right hand-side of
the monitor to silence the alert.
BED PAD PLACEMENT
• Place the pad under the residents shoulders to be
alerted when the resident sits up in the bed (fig 1).
• Place the pad under the resident’s lower back/buttocks to
be alerted when the resident vacates the bed (fig 2).
• Place the monitor out of reach of the resident. The suitable
mounting locations include: the back of headboard, the wall
or under the bed. Make sure the resident cannot tamper with
the monitor.
Chair Pad System
TEST THE SYSTEM
• Plug the pad into the monitor and apply pressure to pad to
activate the system. The monitor will beep twice when active.
• The alert will sound when pressure is released from the pad.
Press the “reset” on the right hand-side of the monitor to
silence the alert.
WHEELCHAIR/CHAIR PAD PLACEMENT
• Center the pad on the chair seat.
• Place the monitor out of reach of the resident. The
suitable mounting locations include: the back or handle of
the wheelchair. Make sure the resident cannot tamper with
the monitor.
Troubleshooting
• Check the power supply. If you are using battery power, install a new 9V battery.
Or, if you are using an AC adapter, make sure the adapter is plugged into a power outlet.
• Make sure that the sensor pad is plugged into the fall monitor.
• Check monitor switches:
A. Check that the volume switch in the battery compartment is set correctly
B. Check that the alert delay switch is set correctly.
• If the steps above do not resolve your problem, call your distributor to troubleshoot.
Make sure both the sensor pad and the fall monitor are with you at the time of the call.
Sensor Pad Maintenance
• When using this sensor pad with foam or gel cushions, test the system to make sure it functions properly.
• To reduce the spread of infection, it is recommended to use a new sensor pad for each patient.
• Clean this sensor pad with disinfectant wipes or anti-bacterial cleaners.
• DO NOT fold or crease this sensor pad.
• DO NOT submerge this sensor pad in any liquid cleaning solution.
• DO NOT expose this sensor pad to high temperature sanitizers. Do not sanitize with an autoclave.
CAUTION:
The manufacturer does not claim that this device will prevent falls and/or elopement. Sensor pads and and fall monitors are prone
to difficulties. Be sure cord cannot entangle or choke the user. Test this device before each use. This device is designed to augment caregivers’
comprehensive resident mobility management program and is not a substitute for proper nursing care. The effectiveness of this monitoring
device relies entirely on the direct supervision and immediate response by the caregiver to this monitor’s alarm. Failure to compy with this
warning may result in injury or death. READ THE INSTRUCTIONS AND LEGAL DISCLAIMER.
Warning:
FAILURE TO COMPLY WITH WARNING MAY RESULT IN INJURY OR DEATH. This device is not suitable for all individuals. Other devices
may be required. This device is not a substitute for visual monitoring by caregiver. The manufacturer does not claim that this device will stop
elopement and/or stop falls. This device is designed to augment caregivers’ existing comprehensive resident mobility management program.
Test this device before each use. Read the instructions and Legal Disclaimer before each use.
This
Pad and Fall Monitor System is designed to be installed by the end-user. As such, it is the entire responsibility of the buyer to ensure that
the system is properly installed and tested. Further, the system is not designed to replace good caregiving practices including, but not limited to:
a.) Direct patient supervision
b.) Adequate training for staff personnel for resident fall management and elopement
c.) Testing of the system before each use
WHERE PERMISSIBLE, IT IS THE SELLER’S INTENT TO LIMIT ANY LIABILITY FOR SPECIAL, CONSEQUENTIAL OR INCIDENTAL ECONOMIC
DAMAGES TO REFUND OF THE PURCHASE PRICE OR REPLACEMENT OF THE GOODS. SELLER DISCLAIMS ANY LIABILITY FOR COMPENSATORY
OR OTHER DAMAGES ARISING OUT OF THE USE OF THE GOODS.
Limited warranty
Seller warrants that the goods sold will be free from defects of workmanship and manufacturing, for a period of one year from the date of sale.
Other than this limited warranty, seller makes no express or implied warranties. Except for this limited warranty, NO EXPRESS OR IMPLIED
WARRANTIES ARE GIVEN AND NO AFFIRMATION OF SELLER, BY WORDS OR ACTION, WILL CONSTITUTE A WARRANTY. THE GOODS BEING
PURCHASED ARE BEING SOLD ON A “WITH ALL FAULTS” BASIS. THE ENTIRE RISK AS TO THE QUALITY AND PERFORMANCE OF THE GOODS
IS WITH THE BUYER. SHOULD THE GOODS PROVE DEFECTIVE FOLLOWING THEIR PURCHASE, THE BUYER, AND NOT THE MANUFACTURER,
DISTRIBUTOR, OR RETAILER, ASSUMES THE ENTIRE COST OF ALL NECESSARY INSTALLATION, SERVICING OR REPAIRS.
SMART CAREGIVER CORPORATION
1229 North McDowell Blvd., Petaluma, CA. 94954 | Tel: 800-650-3637
[email protected] | www.smartcaregivercorp.com
Need Set-up Help? (800) 650-3637
Basic Fall Monitor
1
2
Ins
tal
lat
ion
an
d U
se
Ins
tru
ctio
ns
SM
AR
T C
AR
EG
IV
ER
C
OR
PO
RA
TIO
N
Bas
ic
Fal
l M
oni
tor
• Safety Auto-Reset
• Adjustable
Volume