caution
• The control solution test should be conducted between
59-104°F (15-40°C) and in a location with a relative humidity
of 10-90%.
• Do not touch the control solution to the sample port on the strip
before seeing the message “Apply a drop of control solution”. The
iGlucose® Blood Glucose Meter performs an internal check that
takes 3 seconds. Touching the control solution to the sample port
before prompted will result in an error message.
• Do not drip the control solution on to the sample port of the
test strip directly as this could flood and contaminate the meter
via the test strip port.
2 7. 6 C
Apply a drop o
f
control solution
Standby Mode