Page 71 of 128
Step 4. The SLE1000 is shipped
with the main battery
disconnected.
Using a screw driver
remove the battery fuse
holder. (Indicated by
arrow)
Step 5. Insert the fuse into the
fuse holder. (T2.0A
250V). (The fuse is
located in the accessory
pack, Part nº: M0457/
2AC/001)
Step 6. Insert the fuse holder and
fuse into the receptacle.
Do not over tighten.
Step 7. Remove the screw and
washer holding the P clip.
Insert the power cable
into the socket and secure
the cable with the P clip.
Allow 75mm of cable to
form the loop.
75mm
Содержание SLE1000
Страница 1: ...SLE1000 User manual Version 2 6 4 software ...
Страница 4: ...Page 4 of 128 This page is intentionally blank ...
Страница 8: ...Page 8 of 128 This page is intentionally blank ...
Страница 9: ...Page 9 of 128 Introduction ...
Страница 12: ...Page 12 of 128 This page is intentionally blank ...
Страница 13: ...Page 13 of 128 SLE1000 Description ...
Страница 17: ...Page 17 of 128 Operational and Clinical Warnings ...
Страница 23: ...Page 23 of 128 User interface ...
Страница 36: ...Page 36 of 128 Push the knob to select ...
Страница 54: ...Page 54 of 128 This page is intentionally blank ...
Страница 55: ...Page 55 of 128 CPAP Flow mode ...
Страница 68: ...Page 68 of 128 This page is intentionally blank ...
Страница 69: ...Page 69 of 128 SLE1000 Set up ...
Страница 79: ...Page 79 of 128 Functional Testing ...
Страница 85: ...Page 85 of 128 Operational Considerations ...
Страница 87: ...Page 87 of 128 Operation ...
Страница 89: ...Page 89 of 128 Troubleshooting Chart ...
Страница 92: ...Page 92 of 128 This page is intentionally blank ...
Страница 93: ...Page 93 of 128 Technical Information ...
Страница 120: ...Page 120 of 128 This page is intentionally blank ...
Страница 121: ...Page 121 of 128 Consumables Accessories ...
Страница 123: ...Page 123 of 128 Glossary ...
Страница 128: ......