3.2 Illustrated installation procedure
(j) Dismantle:
® Wall mounting without brackets,
2.0
-2.4M above ground
@Bracket mounting (optional)
A. Insert the screwdriver
Wall
mounting
�------�
B. Mark the drilled hole
and drill
C. Insert two expandable
dowels and attach the
base into wall by four
screws
D.
Replace the
dismantled parts to
base cover
3 pulse: after detecting 3 pulse, it will send alarm signal.
4. Dip switch function specification
D-PIR-I MW can choose following detection mode.
2
2 pulses: after detecting 2 pulses, it
will
send alarm signal (factory default)
ON 3-PULSE
SET PULSE
OFF 2-PULSE
ON PIROR MW
WORK MODE
OFF PIR AND MW
ON LEDON
ALARM LED
More pulses means less sensitivity, but also can decrease the false alarm.
OFF
D-PIR-I MW apply both PIR and microwave technology, it can choose two working mode below
LED OFF
ALARM MODE
15MINS MODE
SMINS MODE
TESTING MODE
3
4
And mode: when both PIR and microwave detects intruder, it will give alarm signal.
Or mode: when PIR or microwave detect intruder, it will give alarm signal.
NO
OFF
Alarm Led on/off: for special cases, users can set alarm led visible or not according to the
OFF
ON
needs.
OFF OFF
D-PIR-I MW can set two alarm modes as below
5mins mode: after alarm, there is 5mins dormancy, trigger the detector during this time, will not alarm.
15mins mode: after alarm, there is 15mins dormancy, trigger the detector during this time, will not alarm.
Testing mode: after alarm, there is 3s dormancy, trigger the detector during this time, will not alarm. (factory default)
5. Coding method between detector
and panel:
6.Change battery:
Coding setting,
CD Set detector as Normal mode, place the battery and LED will flash seconds.
Set panel as Coding mode.(Panel coding please refer to panel manual),
within 3 seconds when press the configure key of the panel,
a. Wave hands near the front side of D-PIR-I MW, detector will
send a alarm signal to the panel. If the panel sounds a response
then code successfully.
b. Inverse the detector and detector will send alarm signal to the panel,
vibration switch,if the panel sounds a response then code successfully.
® Enter the address code to code with the panel. Set the panel as manual
coding mode and enter the 9-digit address code. This will be a higher
Probability of coding success.
7. Walk test in coverage area:
CD Set as Test Mode to proceed walk-test, pulse count set as 1,2 or 3.
When the voltage of the detector is lower than 4.8V,
the control panel will receive low battery signal.
After that, each 2 hours, it will send low battery.
The user should buy the same battery and change.
Place new lithium
battery
® Walk across the far edge of coverage area at the speed of 1 step/second(about0.75m/s)
The LED will flash for seconds then alarm (as shown in the right figure) .
@ Do walk-test in opposite direction to confirm the boundary of both sides, Make
sure the detection centre pointing to the centre of protected area.
@ Make sure the detection centre at the proper place. Should properly adjust the
detection area if you can not get an ideal detection area.
@After adjust the detection angle , should redo the walk test as above.
® Please change TEST mode to NORMAL mode after the Walk-test .
8.Customer service
Our products are very reliable, but for some special reasons, the working performance will be limited in certain range. We here list
some cases as follows,
CD. The voltage of control panel is not stable;
®· Low-voltage of the detector.
For any help please contact with our company and your could visit our website for more information.,
A
Warning: We are not responsible for the problem caused by improper operation by users!