SKF TKTL 21 Скачать руководство пользователя страница 14

14 EN

10. Notes

How does Infrared work?

Infrared thermometers measure the surface temperature of an 
object. The unit’s optics sense emitted, reflected, and transmitted 
energy, which is collected and focused onto a detector.
The unit’s electronics translate the information into a temperature 
reading, which is displayed on the unit.
The laser is used for aiming purpose only.

•  

Field of View

Make sure the target is larger than the unit’s spot size.
The smaller the target, the closer you should be.
When accuracy is critical, make sure the target is at least twice as 
large as the spot size.

•  

Distance & Spot Size

As the distance (D) from the object increases, the spot size (S) of the 
area measured by the unit becomes larger.
See figure at the beginning of this manual.

• 

Locating a hot Spot

To find a hot spot aim the thermometer outside the area of interest, 
then scan across with an up and down motion until you locate hot 
spot.

• 

Reminders

 

-  Do not use the unit to measure shiny or polished metal 

surfaces (stainless steel, aluminum, etc.). See Emissivity.

 

-  The unit can not measure through transparent surfaces such 

as glass. It will measure the surface temperature of the glass 
instead.

 

-  Steam, dust, smoke, etc., can prevent accurate measurement 

by obstructing the unit’s optics.

• 

Emissivity

Emissivity is a term used to describe the energy-emitting 
characteristics of materials.
Most (90% of typical applications) organic materials and painted or 
oxidized surfaces have an emissivity of 0.95 (pre-set in the unit).
Inaccurate readings will result from measuring shiny or polished 
metal surfaces. To compensate, cover the surface to be measured 
with masking tape or black paint. Allow time for the tape to reach the 
same temperature as the material underneath it.
Measure the temperature of the tape or painted surface.

Содержание TKTL 21

Страница 1: ...Instructions for use SKF TKTL 21...

Страница 2: ...5 4 Technical data 6 5 Front panel and button description 8 6 Display overview 10 7 Menu overview 11 8 Measurement settings 12 9 Maintenance 13 9 1 Storage cleaning 13 9 2 LCD error messages 13 9 3 B...

Страница 3: ...repair work should be taken care of by an SKF repair shop Here is a table of hazardous substance name and concentration for TKTL 21 Part name Hazardous Substances Lead Mercury Cadmium Hexavalent Chro...

Страница 4: ...of the following Directive s EMC DIRECTIVE 2014 30 EU and are in conformity with the following standards EN 61326 1 2013 EN 61326 2 2 2013 Emission CISPR 11 2009 A1 2010 Class B Immunity IEC 61000 4 2...

Страница 5: ...the measure target with Lens 7 and press Meas key 6 to display the surface temperature 2 The Distance Spot is 30 1 The eight laser points are the reference for the target spot size Please make sure th...

Страница 6: ...me 1 000 msec 90 Optics Sense emitted reflected and transmitted energy Display Color LCD User selectable backlit display No permanently ON Displayed resolution 0 1 C F from 83 2 999 9 otherwise 1 C F...

Страница 7: ...kaline type IEC LR03 Battery lifetime Min 30 hours continuous use without laser Min 3 hours continuous use with laser and white LED Switch off IR mode automatic after 60 seconds after trigger is relea...

Страница 8: ...8 EN 5 Front panel and button description 6 5 1 2 3 4 Item Description 1 LCD 2 Mode key 3 Down key 4 Up key 5 Thermocouple socket 6 Meas key...

Страница 9: ...9 EN 7 9 10 8 Item Description 7 Infrared lens 8 LED 9 Multi beams 10 Battery cover...

Страница 10: ...10 EN 6 Display overview Default screen Multi Beams spicify the apprximate measurement area for better targetting...

Страница 11: ...3 to change the High Alarm HAL or Lo Alarm LAL then press Meas key 6 to confirm it When the reading is outside the High Alarm HAL or Lo Alarm LAL limit The High or Low icon will flash and you will hea...

Страница 12: ...sferred In MAX MIN mode Hold on the Meas key 6 The Bar display indicates the measuring temperature The bar shows RED color when the reading is close to maximum value and shows BLUE when close to minim...

Страница 13: ...the working room temperature Error 5 9 for all other error messages it is necessary to reset the thermometer To reset it turn the instrument off remove the battery and wait for a minimum of one minut...

Страница 14: ...ot aim the thermometer outside the area of interest then scan across with an up and down motion until you locate hot spot Reminders Do not use the unit to measure shiny or polished metal surfaces stai...

Страница 15: ...80 to 0 95 Water 0 92 to 0 96 Lacquer matt 0 97 Ice 0 96 to 0 98 Rubber black 0 94 Snow 0 83 Plastic 0 85 to 0 95 Glass 0 90 to 0 95 Timber 0 90 Ceramic 0 90 to 0 94 Paper 0 70 to 0 94 Marble 0 94 Chr...

Страница 16: ...ight of the publisher and may not be reproduced even extracts unless prior written permission is granted Every care has been taken to ensure the accuracy of the information contained in this publicati...

Отзывы: