background image

Instructions for use

SKF TKTL 21

Содержание TKTL 21

Страница 1: ...Instructions for use SKF TKTL 21...

Страница 2: ...5 4 Technical data 6 5 Front panel and button description 8 6 Display overview 10 7 Menu overview 11 8 Measurement settings 12 9 Maintenance 13 9 1 Storage cleaning 13 9 2 LCD error messages 13 9 3 B...

Страница 3: ...repair work should be taken care of by an SKF repair shop Here is a table of hazardous substance name and concentration for TKTL 21 Part name Hazardous Substances Lead Mercury Cadmium Hexavalent Chro...

Страница 4: ...of the following Directive s EMC DIRECTIVE 2014 30 EU and are in conformity with the following standards EN 61326 1 2013 EN 61326 2 2 2013 Emission CISPR 11 2009 A1 2010 Class B Immunity IEC 61000 4 2...

Страница 5: ...the measure target with Lens 7 and press Meas key 6 to display the surface temperature 2 The Distance Spot is 30 1 The eight laser points are the reference for the target spot size Please make sure th...

Страница 6: ...me 1 000 msec 90 Optics Sense emitted reflected and transmitted energy Display Color LCD User selectable backlit display No permanently ON Displayed resolution 0 1 C F from 83 2 999 9 otherwise 1 C F...

Страница 7: ...kaline type IEC LR03 Battery lifetime Min 30 hours continuous use without laser Min 3 hours continuous use with laser and white LED Switch off IR mode automatic after 60 seconds after trigger is relea...

Страница 8: ...8 EN 5 Front panel and button description 6 5 1 2 3 4 Item Description 1 LCD 2 Mode key 3 Down key 4 Up key 5 Thermocouple socket 6 Meas key...

Страница 9: ...9 EN 7 9 10 8 Item Description 7 Infrared lens 8 LED 9 Multi beams 10 Battery cover...

Страница 10: ...10 EN 6 Display overview Default screen Multi Beams spicify the apprximate measurement area for better targetting...

Страница 11: ...3 to change the High Alarm HAL or Lo Alarm LAL then press Meas key 6 to confirm it When the reading is outside the High Alarm HAL or Lo Alarm LAL limit The High or Low icon will flash and you will hea...

Страница 12: ...sferred In MAX MIN mode Hold on the Meas key 6 The Bar display indicates the measuring temperature The bar shows RED color when the reading is close to maximum value and shows BLUE when close to minim...

Страница 13: ...the working room temperature Error 5 9 for all other error messages it is necessary to reset the thermometer To reset it turn the instrument off remove the battery and wait for a minimum of one minut...

Страница 14: ...ot aim the thermometer outside the area of interest then scan across with an up and down motion until you locate hot spot Reminders Do not use the unit to measure shiny or polished metal surfaces stai...

Страница 15: ...80 to 0 95 Water 0 92 to 0 96 Lacquer matt 0 97 Ice 0 96 to 0 98 Rubber black 0 94 Snow 0 83 Plastic 0 85 to 0 95 Glass 0 90 to 0 95 Timber 0 90 Ceramic 0 90 to 0 94 Paper 0 70 to 0 94 Marble 0 94 Chr...

Страница 16: ...ight of the publisher and may not be reproduced even extracts unless prior written permission is granted Every care has been taken to ensure the accuracy of the information contained in this publicati...

Отзывы: