1.  Flush the bag at least 3 times 

with puri

fi

 ed air or nitrogen 

before use.

2. To 

fi

 ll a bag, connect PTFE 

tubing from the exhaust port 
of an air sample pump to the 
stem on the bag 

fi

 

tt

 ing.

3.  To open the valve, turn the knurled disc on the stem 

counterclockwise a maximum of 2 revolutions. Turn on the 
pump and sample as usual (

see Proper Bag In

fl

 ation

).

Do not use sample bags to sample compounds with 
boiling points > 249.8 F (121 C).

4.  When sampling is complete, turn o

  the pump and close 

the valve by turning the knurled disc clockwise until snug.

The outside diameter of the syringe needle should not 
be larger than 12 gauge (0.109 inch).

Do not allow the syringe needle to puncture the bag 
material when piercing the septum.

5.  To withdraw samples using a needle and 

syringe, carefully insert the needle into the 
stem and pierce the septum centered in the 
base of the syringe port.

Proper Bag In

fl

 ation

Avoid 

fi

 lling any bag to more than 80% of its maximum volume. At the end of the 

sampling period, turn o

   the pump and close the valve by turning the knurled disc 

clockwise until snug. 

Incorrect 

In

fl

 ation

Correct 

In

fl

 ation