Charging Train with single PowerFlex charger

Power 

cable

Charger

Getting Started

Install the Battery Pack

To enhance battery life, SKC ships battery packs separately from the pump. Once installed, completely charge the battery pack before 
operating the pump.

1.  Remove the two screws that secure the battery pack and loosen 

the two case screws above and below the belt clip.

2.  Carefully slide the battery pack out from under the belt clip. 

Ensure that the battery is kept level.

3.  Slip the front edge of the new battery pack under the belt clip 

and position the battery pack to engage the grooves in the case.

4.  Slide the battery pack toward the pump until it is flush with the 

pump case on all sides.

5.  Reinstall two battery screws and tighten the case screws loosened 

in Step 1.

6.  Charge the battery completely. 

See Charge the Battery Pack. 

Note:

 

For optimum charge, ensure the pump is not running during charging.

Charge the Battery Pack

1.  Activate the charger. 

Follow PowerFlex Operating Instructions 37783.

  For optimum charge, ensure that the pump is not running.

2.  Insert the charging cable connector into the charging port on the charger.

3.  Insert the charging plug end of the cable into the jack on the back of the pump’s 

battery pack

 (Figure 1).

4.  Charge the battery completely (the LED at the port connected to the pump is a 

steady green) before using the pump.

5.  Disconnect the charging cable from the back of the pump’s battery pack.

Notes and Cautions

•  To comply with intrinsic safety regulations, do not charge or operate the pump from the charger in hazardous locations.

•  Using a non-approved charger voids any warranty.

•  Using a repaired or rebuilt battery pack voids any warranty and the UL Listing for intrinsic safety.

•  Using any device other than the approved battery pack to power the pump voids the UL Listing for intrinsic safety and any 

warranty.

•  Ensure proper orientation of charging cable before plugging it into the charging jack. Improper orientation/contact will short-circuit 

the battery and voids any warranty.

•  Short-circuiting the battery pack will render it immediately inoperative.

•  Failure to follow warnings and cautions voids any warranty.

•  The battery pack may be kept on the SKC-approved charger for an indefinite time.

For more information on SKC pump battery packs, visit www.skcinc.com/catalog/pdf/instructions/1756.pdf.

Charging 

cable

Back of pump 

with battery pack 

sliding out

Belt clip

Case screw 

(not shown)

Battery pack 

screws

Case screw 

Case grooves

Battery 

connector

Содержание 224-44XR

Страница 1: ...tion The 44XR Universal Sample Pump Figure 1 is a constant flow from 5 to 5000 ml min air sampler suitable for a broad range of applications It is ideal for industrial hygiene studies as well as environmental testing Checking Pump Kit Contents Use the table below to verify that you received all items associated with the Cat No ordered If you are missing items contact SKC at 800 752 8472 U S only o...

Страница 2: ... on the charger 3 Insert the charging plug end of the cable into the jack on the back of the pump s battery pack Figure 1 4 Charge the battery completely the LED at the port connected to the pump is a steady green before using the pump 5 Disconnect the charging cable from the back of the pump s battery pack Notes and Cautions To comply with intrinsic safety regulations do not charge or operate the...

Страница 3: ...ump operation with Battery Eliminator The Battery Eliminator is an accessory that converts alternating current AC to direct current DC from which the pump can be operated for extended runs The Battery Eliminator should be used in non hazardous locations only See Accessories 1 Remove the battery pack from the pump see Replace the Battery Pack The Battery Eliminator is comprised of two pieces a wall...

Страница 4: ...p has run for five minutes before calibrating Leave the pump running 2 Prepare the calibrator see calibrator instructions Using flexible tubing connect the calibrator outlet suction port to the sampling medium inlet 3 Using 1 4 inch Tygon tubing connect the sampling medium outlet to the pump inlet Figure 2 4 Set the flow rate using the flow adjustment screw Figure 1 until the calibrator indicates ...

Страница 5: ...line trap between the pump and the impinger Mount the impinger and trap to the face of the sampler using the accessory mounting screws Figure 1 or place them in a holster at the worker s waist Failure to use an appropriate in line trap during impinger sampling voids any warranty c When using the pump for pressure applications such as bag sampling thread the exhaust port fitting supplied with the p...

Страница 6: ...ng Leave the pump running 2 Ensure that the regulator has been activated for low flow and the pump flow rate is set at 1 5 L min 3 Connect the single adjustable tube holder to the pump inlet See calibration train 4 Insert an opened sorbent tube into the rubber sleeve Figure 3 of the low flow tube holder with the arrow on the tube pointing toward the tube holder 5 Prepare the calibrator See calibra...

Страница 7: ...used to set the flow with a new unexposed sorbent tube for sample collection 2 Place the appropriate size protective tube cover over the tube and screw it into place on the tube holder For personal sampling clip the tube holder to the worker in the breathing zone See right 3 Start sampling by turning on the pump using the on off switch Figure 2 and record the start time 4 At the end of the samplin...

Страница 8: ... running 2 Ensure that the regulator has been activated for low flow and the pump flow rate is set at 1 5 L min 3 Calculate the sum of all tube flow rates The maximum flow rate through any one tube is 500 ml min a If the sum is 1000 ml min set the pump flow rate to 1 5 L min b If the sum is 1000 ml min multiply that number by 0 15 and total the two numbers Set the pump flow rate for the resulting ...

Страница 9: ... use the pump with the Battery Eliminator in hazardous locations UL Listing for intrinsic safety is not in effect during pump operation with Battery Eliminator Use of any device other than the approved battery pack or Battery Eliminator to power the pump voids any warranty Charge pump battery completely before calibration and sampling Calibrate verify pump flow rate before and after each sampling ...

Страница 10: ... use See Charge the Battery Pack For more information on SKC pump batteries visit http www skcinc com instructions 1756 pdf To comply with intrinsic safety regulations battery packs should not be charged in hazardous locations Ensure proper orientation of charging cable before plugging it into the charging jack Improper orientation contact will short circuit the battery and voids any warranty Shor...

Страница 11: ...ry Maintenance PowerFlex Charging System for SKC Personal Pumps 5 Stations 100 240 V 223 1000 Single 100 240 V 223 2000 PowerFlex Pump Cable for Universal XR models 223 1002 Battery Eliminator connects pump to line power for extended sampling 115 V 230 V 223 325 223 325B Pump Accessories Screwdriver Set included with pump 224 11 Protective Nylon Pouch with belt and shoulder strap Black 224 87 Red ...

Страница 12: ...P22417K Pulsation Dampener Assembly 2 P22417W Bottom Plate Assembly P22433L Flowmeter Assembly Parts Not Shown in Illustration P21251 Half Stack includes pump body valve plates diaphragm yoke gaskets and O rings P2243201 Charging Jack pk 5 P22433C Tamper resistant Cover P22433ES External Screw Kit Replacement Filters P22409 Replacement Filter Set 3 filters 3 O rings P2240901 Filters only pk 10 P22...

Страница 13: ...in rotameter with 250 ml division scale marked at 1 2 3 4 and 5 L min Power Supply 6 V plug in NiMH battery pack rechargeable 3 5 Ah capacity A Battery Eliminator is available see Accessories use voids the UL Listing for intrinsic safety Charging Time varies with capacity and level of discharge 6 to 8 5 hrs with PowerFlex charger Intrinsic Safety UL Listed for Class I Division 1 and 2 Groups A B C...

Страница 14: ...0 14 Polyvinyl Chloride PVC filter 5 0 µm pore size Flow Rate L min Filter Diameter 37 mm 25 mm 2 47 41 2 5 38 33 3 35 30 3 5 26 27 4 22 25 4 5 21 23 Table 1 Pump Run Time in Hours with NiMH Battery Appendix Cont Typical Run Times Table 1 contains the typical run times achieved when using a fully charged nickel metal hydride NiMH battery pack Data is sorted by type of sample media All run times ar...

Отзывы: