4 Installation
64 08 137 D3610
D3610.201.02.09.02 11.2017
53
4.10
Connecting the sensor
A calibration file is saved on the XIOS XG Select and XIOS XG
Supreme sensors which is transferred to the PC the first time that the
sensor is used.
ü
The WiFi interface is configured for the practice network.
1.
Start SIDEXIS XG / SIDEXIS 4.
2.
Plug the connector of the sensor into the WiFi interface.
3.
Register a patient in SIDEXIS XG / SIDEXIS 4 and establish
readiness for intraoral exposures.
Ä
A sensor calibration file is installed automatically if you are
using the XIOS XG Select or XIOS XG Supreme sensor on this
PC for the first time. If this does not occur and no exposure
readiness is established, disconnect the sensor from the WiFi
interface and plug the cable in again.
4.11
Complete test exposures/acceptance tests
A test exposure must be completed after installing a new WiFi interface
or a new sensor.
See operating instructions for the sensor.
In Germany (Austria and Switzerland), an acceptance test should be
performed by an authorized technician of a specialist dealer.
Perform test exposures on a test specimen, see Consumables and
spare parts [→ 108]. Do not perform test exposures on a patient.