5 Care, cleaning and maintenance by the user
Sirona Dental Systems GmbH
5.3 Instruments and instrument hoses
Operating Instructions Simulation Unit SIM MODULAR
5.3
Instruments and instrument hoses
5.3.1
Rinse water lines
Microorganisms can grow in the water lines of the simulation unit. Use a
large amount of water to rinse the lines prior to the use.
5.3.2
Purge the water lines (purge function)
with 3-way syringe Standard and SPRAYVIT E
To reduce the amount of germs, the water lines of the water-carrying
instruments of the dentist element and the Standard or SPRAYVIT E 3-
way syringe of the dentist or assistant element can be purged with water.
For the purge function, individual instruments are removed from the
holder and held over the cuspidor for purging. The water lines of all
removed instruments are then purged consecutively. Press the water key
of the 3-way syringe for purging.
With the SIM MODULAR simulation unit, the purge and sanitizing function
is optional.
With the SIM MODULAR simulation unit, it is also possible to
automatically purge the water lines, see "Purging water lines
automatically (autopurge function)"..
Preparation
The following preparations should be made before you begin to purge the
water lines.
1. Set all of the instruments to be purged to their maximum water flow
rate. The water regulators under the instrument holder must be
turned counterclockwise.
2. Put all instruments in place.
IMPORTANT
The water flow rate to the instruments is not checked by the simulation
unit.
➢ Make sure that the maximum water flow rate is set for all
instruments to be purged.