11 Initial startup
Sirona Dental Systems GmbH
11.10 Service routines for startup
Installation Manual ORTHOPHOS XG 3D / Ceph
11.10.4.8
S018: Test step 9 (occlusal bite block only)
Adjustment of the AD acceptance angle values and the zero point
In the adjustment of the AD acceptance angle values, three AD values
must be adjusted and saved:
●
ZP (zero point):
Zero point; applied whenever no bite block is plugged
in.
●
U (Upper):
Upper value; applied whenever the bite block is plugged
in and the bite block plate is deflected all the way up.
●
L (Lower):
Lower value; applied whenever the bite block is plugged in
and the bite block plate is deflected all the way down.
➢ Call service routine S018.9 [ → 125].
The data type intended for the next storage function is displayed
in selection field 1.
The display of patient symbol key 1 (G) indicates whether or not
the bite block is connected (see table).
Symbol on the control
panel
Status
Meaning
Patient symbol key 1 (G) lit
Occlusal bite block
connected
not lit
Occlusal bite block not
connected
Selection field
Display
Meaning
1
ZP
Zero point
2
U
Upper value
3
L
Lower value