11 Initial startup
Sirona Dental Systems GmbH
11.8 Checking the data paths
Installation Manual ORTHOPHOS XG 3D / Ceph
11.8.2
Creating a 3D test image
✔
The
"Test exposures/test images"
1. In the structure tree, under
"Test exposures 3D"
, select the
"Test
image Sensor"
element.
The menu for creating the test image for the 3D sensor appears
in the action window.
2. Click on the
"Image acquisition"
button to create the test image.
The exposure window opens in SIDEXIS XG. It indicates the
current status of exposure readiness.
3. Press the release button. Press and hold down the button until the
exposure is complete, the preview image is displayed in the exposure
window, and the acoustic signal that indicates the end of the
exposure (double beep) sounds (if it has been configured).
The test image appears in the exposure window.