11
Appendix
Sirona Dental Systems GmbH
11.1
Service routines (for installation)
Installation Instructions ORTHOPHOS XG 3
60 51 416 D 3352
108
D 3352.031.03.18.02
11.1.10Canceling the minimum travel height setting (S018.
6)
1.
Select test step 6 in selection field 2 with the
arrow
keys
and confirm your selection by pressing the
Ser
-
vice key
.
The current height position is displayed in selection
field 1. The LED above the
Memory key
lights up
.
2.
To cancel the limitation for the minimum travel height,
first press the
Memory key
(LED above the R key
lights up), and then the
R key
.
•
Quit service routine S018 (see page 101).
•
Switch the unit OFF and restart it.
After the restart, the previously set minimum travel
height is cancelled.
1.
1
2
3
S018_______6____
➊
➋
2.
➊
R
➋
1
2
3
__200_mm________
R
Содержание orthophos xg 3
Страница 1: ...NMKOMNR båÖäáëÜ kÉï ë çÑW loqelmelp ud P fåëí ää íáçå fåëíêìÅíáçåë T R Prog S kV mA P1 ...
Страница 5: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 5 1 Before you begin ...
Страница 13: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 13 2 Delivery and transport ...
Страница 19: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 19 3 Installation Panoramic X ray unit ...
Страница 45: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 45 4 Electrical connection ...
Страница 49: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 49 5 Installation Remote control ...
Страница 59: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 59 6 Safety checks ...
Страница 65: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 65 7 Initial startup ...
Страница 73: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 73 8 Startup for USA Canada only ...
Страница 83: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 83 9 Checking the system adjustment ...
Страница 91: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 91 10 Final work ...
Страница 97: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 97 11 Appendix ...
Страница 141: ......