2.6 List of available service routines
2.6
2.6
List of available service routines
Routine
Description
Required...
see page
S001
Radiation without rotary movement:
Fixed radiation time of 14.0 s
z
for tube assembly test and testing kV/mA levels
HINWEIS
i
This routine enables the service engineer to release
radiation without having to attain exposure readi-
ness for the entire system.
S002
Radiation without rotary movement:
Max. radiation time selectable
Test step 1:
Radiation from any position;
last primary diaphragm setting
Test step 3:
Radiation from any position;
Primary diaphragm opened fully
Test step 4:
Radiation from any position;
Step filter in beam path;
Primary diaphragm symmetrically
widened to PAN setting
z
for system test, final testing, tube voltage mea-
surement, reproducibility measurement, equiv-
alent dose measurement
S005
General X-ray tube assembly service
Test step 2:
Automatic alignment of preheating
(board DX6) with display of setpoint
values
Test step 4:
Fan test (board DX6)
Test step 5:
Temperature sensor test (board DX6)
z
after error messages or component
replacement
Test step 6 (from V02.22):
Switch off the pulse/pause monitoring
z
only if expressly requested by the customer!
VORSICHT
The execution of this service routine is irrevers-
ible.
The system is operated beyond its specifica-
tions afterwards. This may result in an X-ray tube as-
sembly failure.
This entails a loss of warranty for the X-ray tube
assembly!
The performance of this service function can be
proven by Sirona following a system failure.
Test step 7 (from V02.27):
Switch tube assembly fan to continu-
ous operation
z
only if expressly requested by the customer!
VORSICHT
The execution of this service routine is irrevers-
ibly documented on the tube assembly!
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...