59 38 399 D3352
5 – 34
D3352.076.01.13.02
07.2008
5.6 Service routine S005
Tabs 5
S005.6: Switch off cool-down interval of the single tank
5.6.4
S005: Test step 6
Upon explicit customer request, the monitored pulse/pause ratio can be dis-
abled in exceptional cases. This can be done one time via service routine
S005, test step 6. Re-enabling the pulse/pause monitoring function is
not
possible
, though. Such a one-time disablement can be displayed on the info
screen and in the detail query by entering "PulsePause-Disable".
CAUTION
The execution of this service routine is irreversible. The system is
operated beyond its specifications afterwards. This may result in an
X-ray tube assembly failure. As a result, the system warranty expires.
SIRONA is able to prove this system setting after a system failure.
Selecting test step 6
1.
Select test step 6 in selection field 2 with the
arrow keys
and confirm your
selection by pressing the
Service key
.
Confirming the security access
2.
Confirm the security access by once again selecting the number of the
main routine (5) with the
arrow keys
of selection field 2 and then pressing
the
Service key
.
(
1
2
3
+
-
S005
6
n
(
1
2
3
S005_______6____
XG
Plus
:
Easypad touchscreen
XG 5 / 3 / 3 PPE
: Multipad
o
+
-
(
n
(
n
(
o
1
2
3
-
+
5
n
(
1
2
3
___________5____
XG
Plus
:
Easypad touchscreen
XG 5 / 3 / 3 PPE
: Multipad
(
n
(
n
(
o
(
o
Содержание ORTHOPHOS XG 3 DS
Страница 4: ......
Страница 9: ...ORTHOPHOS XG 1General information...
Страница 12: ...59 38 399 D3352 1 4 D3352 076 01 13 02 07 2008 Tab1...
Страница 59: ...ORTHOPHOS XG 2 Messages...
Страница 124: ...59 38 399 D3352 2 66 D3352 076 01 13 02 07 2008 2 6 List of available service routines Tab 2...
Страница 125: ...ORTHOPHOS XG 3 Troubleshooting...
Страница 152: ...59 38 399 D3352 3 28 D3352 076 01 13 02 07 2008 3 10 Check data paths Generate test images Tab 3...
Страница 153: ...ORTHOPHOS XG 4 Adjustment...
Страница 268: ...59 38 399 D3352 4 116 D3352 076 01 13 02 07 2008 4 6 Resetting the adjustment XGPlus 5 3 3 PPE Tabs 4...
Страница 269: ...ORTHOPHOS XG 5 Service routines...
Страница 433: ...ORTHOPHOS XG 6 Repair...
Страница 436: ...59 38 399 D3352 6 4 D3352 076 01 13 02 07 2008 Tab6...
Страница 530: ...59 38 399 D3352 6 98 D3352 076 01 13 02 07 2008 6 21 Replacing cables Tabs 6...
Страница 531: ...ORTHOPHOS XG 7 Maintenance...
Страница 575: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008 7 45 Tab 7 7 14 Checking the device leakage current 7 14...
Страница 577: ...b 59 38 399 D3352 D3352 076 01 13 02 07 2008...