8
Alarms
NITRASEAL Thermal Sealer is equipped with a signal-
ling system that uses acoustic and visual indications to
alert the operator and simplify the service activity.
In case of alarm both the yellow and green indicators will
flash.
In case of alarms call the Service.
9
Maintenance
The appliance does not require any ordinary mainte-
nance.
To wash the outside of the appliance, use a damp cloth
after making sure that the power supply cable has been
unplugged.
For an extra maintenance call the Service.
Table of Alarms
Alarm code
Cause
1 flash
Break of the probe
2 flashes
Short circuit of the probe
3 flashes after 10 seconds from START
The resistance does not heat up
4 flashes
Fault in the double safety circuit (it appears at the start of
the cycle)