62 14 972 D 3507
D 3507
.
031.01.07
.
02 11.2010
69
Sirona Dental Systems GmbH
Appendix
Installation Instructions HELIODENT
PLUS
Service routines
båÖäáëÜ
6. To save the current service routine, press the Adult key.
7. To discard the settings of the current service routine and quit it, press
the Child key.
8. Finally, switch the unit off and then on again.
9.2.2
Service routine S15
empty
Explanation
Configuration of the safety circuit bypass.
The safety circuit can be bypassed via hardware on DX4 (slide switch S2)
(See section "Setting switch S1 and S2 on DX4 [ ➙ 68]".
● Slide switch S2 "
On
" = safety circuit bypassed
● Slide switch S2 "
Off
" = safety circuit activated
If the setting of slide switch S2 is changed, then the setting of service
routine S15 also must be changed.
● If slide switch S2 is set to "
On
", then "
On
" also must be activated in
service routine S15.
● If slide switch S2 is set to "
Off
", then "
Off
" also must be activated in
service routine S15.
empty
Factory setting
● Slide switch S2 and service routine 15 are set to "
On
" ex works.
empty
Operation
Toggle between
Off
and
On
by pressing the +/- keys.
intern
extern
ON
OFF
S1
S2
S1
Содержание Heliodent plus
Страница 1: ...NNKOMNM ebiflabkqmirp k W b f f HELIODENT PLUS 7mA DC 7mA DC 70kV 60kV S 008...
Страница 2: ......