62 14 972 D 3507
68
D 3507
.
031.01.07
.
02 11.2010
Appendix
Sirona Dental Systems GmbH
Selection of the release button
Installation Instructions HELIODENT
PLUS
9
Appendix
9.1
Selection of the release button
● Slide switch S1 toggles between the release button on control
membrane S4 (
internal
) and an external release button
(
external
). The external release button can be a manual release
(S3) or a release button with remote control (S9).
● When delivered, slide switch S1 is set to "
external
".
9.2
Service routines
9.2.1
Operation
1. Switch the unit on.
2. Press the Film key, the Sensor key and the Bite Wing key
simultaneously.
ª
The service routine "
S01
" is displayed in front of a white
background.
3. Scroll through the list of service routines by pressing the +/- keys.
4. Press the Film key to show the setting of the displayed service routine
and make any changes which then may be necessary.
ª
The setting is displayed.
5. To change the setting, press the +/- keys.
intern
extern
ON
OFF
S1
S2
S1
+
+
+
_