Sirona Dental Systems GmbH
7
Adjusting/calibrating the unit
Service Manual GALILEOS
7.3
Adjustment and calibration via the "Service Functions" menu
båÖäáëÜ
7.3.6
Shading calibration
Shading calibration
✔
The
"Service exposure "
1. Click the
"Shading"
tab.
ª
The corresponding tab card is selected.
2. Click on the
"Image acquisition"
button.
ª
SIDEXIS XG makes the unit ready for exposure [ → 162].
ª
Service routine S010.11 or S010.15 (extended shading
calibration) is displayed on the control panel.
3. Press the R key to move the unit back to the starting position.
4. Press the release button. Press and hold down the button until the
exposure is complete, the preview image is displayed in the exposure
window, and the acoustic signal indicating the end of the exposure
(double beep) sounds (if it has been configured).