Sirona Dental Systems GmbH
6 Installation
Installation Instructions Facescan
6.5 Converting the covers of the X-ray detector
båÖäáëÜ
Cable connections to terminal K1
1. Connect cable L76.BN (brown) to terminal K1.1 (orange).
2. Connect cable L76.WH (white) to terminal K1.2 (orange).
3. Fasten cable L76 as low as possible downwards with the cable clamp
(G).
6.5
Converting the covers of the X-ray detector
1. Loosen the screws (S).
2. Loosen the four screws (R) and remove the X-ray detector cover (V).
IMPORTANT
Only remove the parts mentioned in the text.