B1051067 Rev. A
CDR PlusWire User Guide
20
Immunity Test
IEC 60601 Test
Level
Compliance
Level
Guidance
IEC 61000-4-11
40% U
T
(60% dip in
U
T)
for 5
cycles
70% U
T
(30% dip in
U
T)
for 25
cycles
< 5% U
T
(>95% dip in
U
T)
for 5
sec
40% U
T
(60% dip in
U
T)
for 5
cycles
70% U
T
(30% dip in
U
T)
for 25
cycles
< 5% U
T
(>95% dip in
U
T)
for 5
sec
power supply or battery.
NOTE: U
T
is the AC
mains voltage
prior to application of the test level.
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3A/m
3A/m
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial or hospital environment.
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
Portable and mobile RF
communication equipment should
be used no closer to any part of
CDR PlusWire, including its cables,
than the recommended separation
distance calculated from the
equation applicable to the frequency
of the transmitter.
Recommended separation distance:
P
1.2
d
=
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
P
1.2
d
=
for 80 MHz to 800 MHz
P
3
.
2
d
=
for 800 MHz to 2.5 GHz
Where P is the maximum output
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation in meters
(m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
a
should
be less than the compliance level in
each frequency range.
b
Interference may occur in the
vicinity of equipment marked with
the following symbol.