Lamp Replacement
CAUTION:
When it becomes necessary to replace the
lamp, remove the battery and allow the instrument to
cool down before proceeding.
1. Grasp the main barrel and twist counterclockwise
until it reaches the end of the threads. Carefully lift
the barrel away from the instrument (Fig. 4), being
careful not to touch the lamp as it may be hot.
2. Place instrument vertically on a flat surface. Remove the Spacer Ring.
3. Use a magnetic Phillips screwdriver to loosen and remove the two clamp screws
(Fig. 5). This will prevent the screws from falling down inside the instrument.
Carefully remove the clamps.
4. While holding the lamp assembly in one hand, carefully unplug the lamp power
connector (Fig. 6). Set the expired lamp aside.
5. Unpack the new lamp assembly and plug it into the lamp connector.
CAUTION!
Do Not
touch lamp as this will significantly reduce the life of the bulb.
6. Reinstall both clamps one at a time. Do not fully tighten the first clamp. Position
the second clamp and tighten its screw, and then fully tighten the first clamp.
7. Reinstall the Spacer Ring and screw the main barrel onto the instrument.
8. Reinstall the battery and test the lamp.
Fuse Replacement
•
The fuse access is located on the back left side of unit. Using a small flathead
screwdriver, push down and turn knob 1/4-turn counterclockwise to expose the fuse.
•
Replace with 5mm x 20mm 6.3 amp type slow blow 250V fuse. Reinsert fuse
knob, push down and turn 1/4-turn clockwise until it locks in place.
FIGURE 4
FIGURE 6
FIGURE 5
FIGURE 7
FUSE