2.2
Device configuration
Item
Description
Measurement
Record Mode
Select values to be stored or just displayed.
Sync at Full Hour
Synchronize time intervals to full hours.
Fixed Duration
Measurement ends after a fixed time.
Repeat Mode
If a Fixed Duration is specified an automatic repeat can be enabled.
Repeat Interval
If a Fixed Duration is specified and the Repeat Mode is enabled, an interval must be configured
in which the measurement is repeated. So can be configured for example to measure every day
for 1 hour at 6 o’clock.
Start Time
Predefine a time to start the measurement automatically.
Calibration
Start Calibration
Start the process of calibration.
Calibration Level
Set the calibration level (94.0 dB, 104.0 dB, 114.0 dB, Other).
Reset Calibration
Reset the calibration level to factory value.
Measurement Values
(Select a sound level e.g. L
AFmax
and adjust the following properties.)
Display Location
Select on which graph the value shall be shown.
Display Interval
Select the rate of display refresh (500 ms, . . . , 60 min, End Result only).
Storage Interval
Set the storage interval (No storage, 62.5 ms, . . . , 60 min, End Result only).
Print Result
Decide whether the result shall be printed.
Octave Spectrum
Octave Mode
Select whether 1/3 or full octaves shall be measured.
Display Range
Adjust the scaling of the Y-axis.
Display Weigthing
Available weigthings: Z, A, C
Display Interval
Select the rate of display refresh (125 ms, . . . , 60 min, End Result only).
Storage Weigthing
Available weigthings: Z, A, C
Storage Interval
Set the storage interval (No storage, 125 ms, . . . , 60 min, End Result only).
History Graph
Display Range
Adjust the scaling of the Y-axis.
History Time
Adjust the scaling of the X-axis.
Graph 1
Select which value shall be displayed as yellow.
Graph 2
Select which value shall be displayed as green.
Storage Files
Stored Files
View Results, Delete File
Reset Storage File ID
Resets File ID to the smallest possible number, so that no gaps are filled with new measurements:
Assume stored measurements are labled with IDs 1 to 6. If for example measurements 3, 5, and
6 are deleted and the File ID is reset, the next measurement will be labled with the number 5.
Erase all Storage Files
Clear all Files.
Device
Time
Set time and select 12h or 24h format.
Display
Set Display properties.
Panels
Set, whether Panels should be displayed or not:
Octave Spectrum, History Graph, Results, Measurement Info
Options
Additional configurational settings:
Microphone Connection (Direct connection, cable connection, Weather Protector)
Power (automatic: On by USB, On by Reset, Off by Inactivity)
Info
Provides information about firmware and hardware (see also figure 1.6).
Reset Configuration
Reset to factory settings.
Print Results
Table 2.2: Manual configuration menu of Tango_Plus
SINUS Messtechnik GmbH
13 of 47
Manual Tango_Plus
Содержание Tango Plus
Страница 16: ...2 APPLIANCE Figure 2 6 Configuration in setup tab Manual Tango_Plus 16 of 47 SINUS Messtechnik GmbH...
Страница 34: ...3 INFORMATION FOR TYPE APPROVAL Manual Tango_Plus 34 of 47 SINUS Messtechnik GmbH...
Страница 35: ...3 2 Configuration with weather protection Outdoor Kit SINUS Messtechnik GmbH 35 of 47 Manual Tango_Plus...
Страница 48: ......