Simpson Electric 270 Series 5 Скачать руководство пользователя страница 10

10

**Responds to the average value of an AC current, and is calibrated to indicate
the RMS value of a pure sine wave.
***Per ANSI C39.5 April 1974: “The maximum voltage, with respect to Ground,
which may safely and continuously be applied to the circuit of an Instrument.”

1.9

Definition of Accuracy

The voltage and current accuracy of this Instrument is commonly expressed as a
percent of full scale.  This should not be confused with accuracy of reading (indi-
cation).  For example, +2% of full scale on the 10 volt range allows an error of
±0.20V at any point on the dial.  This means that at full scale, the accuracy of
reading would be ±2%, but at half scale it would be ±4%.  Therefore, it is advan-
tageous to select a range which gives an indication as near as possible to full
scale.

1.10 Safety Considerations

This Operator’s Manual contains special caution and warning symbols which
alert the user to hazardous operating and servicing conditions.  These symbols
appear throughout this publication, where applicable, and are defined on the
inside front cover of this manual under NOTES AND SAFETY INFORMATION.
These instructions must be adhered to in order to ensure the safety of operating
and servicing personnel and to retain the operating conditions of the Instrument.

Table 1-2.  Items and Accessories Furnished with this Instrument

Quantity

Description

Number

      1

Test Lead Set, one red and one
black (48” long) each with combination
probe tip and removable rubber-sleeved
alligator clip.

00043

    *1

1.5 volt, D cell, NEDA 13 F

    *1

9.0 volt Cell, NEDA 1604

      2

1 Amp, 250 V Fuse, Littlefuse Type #312001

      1

2 Amp Fuse, 600 V (Bussman Type BBS or
Littlefuse Type BLS)

      1

Operator’s  Manual

6-111642

*Batteries are standard items replaceable from local retail stores.

Table 1-3.  Additional Accessories

Description

Catalog Number

Deluxe Case

00812

Leatherette Ever-Redy Hard Case, Drop Front

00805

Low-Power Ohm Probe

00414

Model 150-2 Amp-Clamp

00545

23

4.17 Zero Ohms Adjustment

When resistance is measured, the INTERNAL batteries B1 and B2 furnish power
for the circuit.  Since batteries are subject to variation in voltage and internal
resistance the Instrument must be adjusted to zero before measuring a resis-
tance, as follows:

a.

Turn range switch to desired ohms range.

b.

Plug the black test lead into the - COMMON jack and the red test lead into the
+ jack.

c.

Connect ends of test leads together to short the VOM resistance circuit.

d.

Rotate the ZERO OHMS control until pointer indicates zero ohms.  If pointer
cannot be adjusted to zero, one or both of the batteries must be replaced.
(For battery replacement refer to Section 5.)

e.

Disconnect shorted ends of test leads.

4.18 Measuring Resistance

Voltage applied to a resistance range will cause reading errors if low and dam-
age the Instrument if high.  When making in-circuit measurements, make certain
that the circuit is completely de-energized before making connections to it.

When making in-circuit measurements, circuit paths in parallel with the resis-
tance being measured may cause reading errors.  Check circuit diagrams for the
presence of such components before assuming that the reading obtained is
correct.

a.

Before measuring resistance in the
circuit make sure the power is off in
the circuit being tested and all capaci-
tors discharged.

b.

Disconnect the shunted component
from the circuit before measuring its
resistance.  Then set the range switch
to one of the resistance range posi-
tions as follows (Figure 4-8):
1.

Use R x 1 for resistance read-
ings from 0 to 200

.

2.

Use R x 100 for resistance read-
ings from 200 to 20,000

.

3.

Use R x 10,000 for resistance
readings above 20,000

.

c.

Set the function switch at either - DC
or + DC position. Operation is the
same in either position except if semi-
conductors are in the circuit.  (See
paragraph 4.19)  Adjust ZERO
OHMS control for each resistance
range.

d.

Observe the reading on the OHMS
scale at the top of the dial.

-D.C.

+D.C.

WARNING

READ MANUAL BEFORE USE

COMMON

ALL TERMINALS 1KV. MAX

270

RX1

RX100

RX10,000

Figure 4-8. Jacks and Switch
Positions for Measuring
Resistance

Содержание 270 Series 5

Страница 1: ...liammeter INSTRUCTION MANUAL 32 SIMPSON ELECTRIC COMPANY 520 Simpson Avenue Lac du Flambeau WI 54538 0099 715 588 3311 FAX 715 588 3326 Printed in U S A Part No 06 111642 Edition 14 05 07 Visit us on the web at www simpsonelectric com ...

Страница 2: ...red by it to be free from defects in material and workmanship under normal use and service its obligation under this warranty being limited to making good at its factory or other article of equipment which shall within one 1 year after delivery of such instrument or other article of equipment to the original purchaser be returned intact to it or to one of its authorized service centers with transp...

Страница 3: ...3 NOTES 30 NOTES ...

Страница 4: ... DC Voltage Measurement 0 2 5 Through 0 500V Range 16 4 8 DC Voltage Measurement 0 1000v Range 16 4 9 AC Voltage Measurement 0 2 5 Through 0 500V Range 17 4 10 AC Voltage Measurement 0 1000V Range 18 4 11 Output Voltage Measurement 19 4 12 Decibel Measurement 20 to 50 dB 20 4 13 Direct Current Measurement 21 4 14 Direct Current Measurement 0 50 µA Range 21 4 15 Direct Current Measurement 0 1mA thr...

Страница 5: ...5 112004 F1 Fuse 1 Amp 250V 3AG 1 1 4 x 1 4 Quick acting Littlefuse Type 312001 1 112507 F2 Fuse 2A 600V Bussman Type BBS or 5 119056 Littlefuse Type BLS For Function Switch 1 115789 Knobs For Zero Ohms Adjust 1 115790 For Range Switch 10 864352 Case Assembly Phenolic Complete including Handleless Battery Compartment Front Cover 10 864996 Cover Assembly 10 861724 Rubber Bumper Plug 5 115039 270 5 ...

Страница 6: ...ned in this manual and should be read and understood thoroughly before ordering replacement parts for the Instrument 6 1 Removing the Instrument Case To open the Instrument panel first de energize and disconnect the test leads and remove the cover from the battery compartment Remove the four case fastening screws and lift the case from the panel When replacing the case make certain that the wires ...

Страница 7: ...use may be burned out Refer to paragraph 5 3 for fuse replacement A 1 Amp spare fuse is furnished with each Instrument Fuses are located in the battery and fuse compartment NOTE The 0 10A range is not fused The shunt connects directly to the 10A and 10A jacks d Two diodes connected across the meter movement afford protection to the movement by circumventing excessive current around the movement in...

Страница 8: ...st probe s to extend the Instrument AC or DC voltage range The test leads and the insulated alligator clips are rated for the voltage and current ranges available on the 270 Maximum ratings are 1000 V AC or DC or 10 DC amperes These ratings for the test leads must not be exceeded in any application 1 8 Technical Data Table 1 1 Technical Data 1 DC VOLTS Ranges 0 1 2 5 10 50 250 500 1000V Sensitivit...

Страница 9: ... increasing values of resistance e To determine the actual resistance value multiply the reading by the factor at the switch position K on the OHMS scale equals one thousand 4 19 Resistance Measurement of Semiconductors Make sure that the OHMS range being used will not damage any of the semicon ductors refer to Table 1 1 item 8 Section 1 If there is a forward and backward resistance such as in dio...

Страница 10: ... 4 17 Zero Ohms Adjustment When resistance is measured the INTERNAL batteries B1 and B2 furnish power for the circuit Since batteries are subject to variation in voltage and internal resistance the Instrument must be adjusted to zero before measuring a resis tance as follows a Turn range switch to desired ohms range b Plug the black test lead into the COMMON jack and the red test lead into the jac...

Страница 11: ...nge positions marked 1 mA 10 mA 100 mA or 500 mA d Turn OFF power Discharge any ca pacitors in circuit e Open the grounded side of the circuit in which the current is being mea sured Connect the VOM in series with the circuit Connect the red test lead to the positive side and the black test lead to the negative side f Turn the power on Read the current in milliamperes on the black DC scale For the...

Страница 12: ...rence level subtract 7 dB from the reading obtained on the 270 5 4 13 Direct Current Measurement a Do not change the setting of the range or change the function switches while circuit under measurement is energized b Never disconnect test leads from circuit under measurement while the circuit is energized c Always turn the circuit power off and discharge all the capacitors before setting the switc...

Страница 13: ...tion often specially designed instruments operated by specially trained personnel who use appropriate protective equipment and procedures 4 1 Safety Precautions The following precautions are commonly recognized safe practices to adopt and specific hazards to avoid and are not implied to be sufficient to safeguard un trained personnel in all circumstances Neither is this manual a substitute for tec...

Страница 14: ...voltage must not exceed the Instrument s rated maximum circuit to ground voltage n Do not measure where the air may contain explosive concentrations of gas or dust such as in mines grain elevators gasoline stations or in the pres ence of charging batteries until determined to be safe by qualified personnel Note that even metallic dusts can be explosive o No general purpose VOM is to be used to mak...

Страница 15: ...t the function switch at DC Figure 4 1 b Plug the black test lead into the COMMON jack and the red test lead into the 1V jack c Set the range switch at 1V COMMON position with 2 5V d Connect the black test lead to the negative side of the circuit being measured and the red test lead to the positive side of the circuit e Read the voltage on the black scale marked DC and use the figures marked 18 e ...

Страница 16: ...WARNING READ MANUAL BEFORE USE 270 17 charged e Connect the black test lead to the negative side of the circuit being measured and the red test lead to the positive side of the circuit f Turn on power in circuit being measured g Read the voltage using the 0 10 figures on the black scale marked DC Multiply the reading by 100 Turn off power to the circuit and wait until the meter indicates zero befo...

Отзывы: