background image

12

12

X-Talk Test

X-TALK is shorthand for Cross-Talk. Cross-Talk is interference between two 
signals, such as IDNet/MAPNET channels. The X-TALK test measures the 
impedance between the POSITIVE lead of two IDNet channels. Coupling can 
be caused by capacitive, resistive or inductive impedance between two 
electrical signals. The greater the coupling between two electrical signals is, 
the greater the likelihood of communications errors.

To initiate the X-TALK Test, navigate to the 
X-TALK screen. Connect the wires as 
shown in the table. Press ENT. The test 
results are displayed when the XTALK Test 
is complete. To test a different cable pair, 
move the Green wire to the +IDNet of the 
next channel, press menu, then when X-
TALK Test is displayed, press ENT. The X-TALK test may be conducted on 
two IDNet channels (connect IDNet+ as described in the table). It may also 
be used between RUI and IDNet, or IDNet and IDNAC.

To Test

Connect

IDNet

RED lead to +IDNet wiring of Channel "A"

BLACK lead is not connected to field wiring

GREEN lead to +IDNet wiring of Channel "B"

Test Results are displayed and indicate the severity of the coupling between 
the channels

LOW Coupling 

- Best Result. Indicates no cross-talk between the channels 

tested

MODERATE Coupling 

- Indicates that coupling is adequate to cause some 

communications errors. The level of cross-talk would usually not be visible at 
the front panel, because the panel re-tries when a communications error takes 
place. If the re-try is successful, the panel will ignore the comms error and will 
not indicate any fault.

HIGH Coupling 

- Indicates that coupling between the two IDNet channels is 

too severe for reliable communications

ENT

X-TALK Test Complete

High Coupling

Failure

Press MENU to exit

Содержание TrueSTART II

Страница 1: ...verify no external voltage source is present before connecting this instrument Use up down arrow and keys to move between menu choices To Navigate Menus Press ENT ket to select a menu choice Press ME...

Страница 2: ...t Shorts 9 IDNAC 10 Power Option 11 X Talk Test 12 Test IDC zone 13 Test NAC Signal 14 Saving IDNet and IDNAC Test Results 15 Recalling Test Results 15 Upload to PC 15 System Settings 16 Backlight Tim...

Страница 3: ...devices are correctly installed before connecting the wires to the FACP Power Sources The TSIT operates from multiple power sources Lithium Ion Battery AC Adapter Any 24V DC source using the Auxiliary...

Страница 4: ...xtend the reach of the AC adapter Provide power via alligator clips to ZAMs and other 24V powered devices when used with the AC adapter Power the tester from an FACP The output from the AC adapter wil...

Страница 5: ...C addressable devices Simplex NACs and Simplex IDCs Connect and test only these types of circuits Before connecting the tester to any circuit for testing verify that the circuit is not connected to an...

Страница 6: ...und such as building steel To detect every earth fault you must test the positive lead and the negative lead of a circuit independently If there are earth faults on both positive and negative leads it...

Страница 7: ...one repeating the process To test an IDNet channel with IDNet Isolators or TrueAlarm Isolator Bases Earth Fault Search must also be run on the IDNet channel IDNet The user should have a list of IDNet...

Страница 8: ...ompare it to the devices that are supposed to be installed Press ENT to review the results of the scan Use the up down arrows to scroll through the menu choices The choices are to review Totals review...

Страница 9: ...in repeating the steps outlined earlier in this document If you just wish to run the Duplicate Address test scroll through the menu choices until that selection is displayed and press ENT The IDNet Te...

Страница 10: ...rom the TSIT The TSIT will not power the appliances into alarm mode At the completion of the scan review the summary of devices detected and compare to the list of devices that are supposed to be inst...

Страница 11: ...evice It is useful to also take note of any missing devices before going out to correct duplicates IDNAC Tests In the example above two devices were set to address 003 and let s assume a device was mi...

Страница 12: ...may be conducted on two IDNet channels connect IDNet as described in the table It may also be used between RUI and IDNet or IDNet and IDNAC To Test Connect IDNet RED lead to IDNet wiring of Channel A...

Страница 13: ...d by the electromagnetic field of a signal interfering with the electrical signal on an adjacent conductor Inductive coupling interference is worse when a current changes rapidly in amplitude Signals...

Страница 14: ...ng Test NAC Signal Test conventional Notification Appliance Circuits with this menu selection At the top level menu navigate to Test NAC signal with the up down arrow keys and press ENT to select the...

Страница 15: ...will surround the slot number and the or keys are used to select the desired slot number 1 60 Press ENT to save to this new slot Note that the TSIT allows you to override a previously filled memory sl...

Страница 16: ...t be erased System Settings The System Settings menu option includes menu choices that allow you to view information clear memory slots and set TSIT options System Settings Backlight Timer Battery Sel...

Страница 17: ...or keys to select the slot to erase or select Clear All to erase all memory slots The ENT key must be pressed twice to erase the memory The second key press is a confirmation of the requested operatio...

Страница 18: ...dard base SPHOTO TrueAlarm photoelectric sensor sounder base RPHOTO TrueAlarm photoelectric sensor relay base IPHOTO TrueAlarm photoelectric sensor isolator base QPHOTO TrueAlarm quick connect photoel...

Страница 19: ...ted CO SOUNDER BASE TrueAlarm CO sensor base with sounder no head inserted UNKNOWN DEVICE Unrecognized device type Type Description INVALID ADDRESS A device was found at one of the following addresses...

Страница 20: ...sual appliance A V Audio Visual appliance ISO Isolator Type Description DUPLICATE 2 or more device set to the same address NO ANSWER A device that previously communicated no longer does OPEN Other por...

Страница 21: ...0 Candela 75Cd 75 Candela 75WP Cd 75 Weatherproof Candela 110Cd 110 Candela 135Cd 135 Candela 185Cd 185 Candela FACP Cd Candela configured by FACP Lens Color Clear Clear lens cover Color Colored lens...

Страница 22: ...nstaller will automatically start If it does not automatically start use Windows Explorer to open the CD folder and double click on the file setup exe You must accept the license agreement to install...

Страница 23: ...ordered from stock SKU Description TSI 24V TSI TSIT 24 V wall wart TSI BATTERY TSI TSIT 24 V battery TSI CHARGER TSI TSIT battery charger TSI PROTECT TSI TSIT case protector TSI TLEADS TSI TSIT test...

Страница 24: ...specifications and other information shown were current as of document revision date and are subject to change without notice TYCO SIMPLEX and the product names listed in this material are marks and o...

Отзывы: