Zero-Point Adjustment: 

If there is no pressure applied to the pressure ports the device will display 0. If  

there is a permanent deviation (and device is operated under steady conditions), a 

permanent zero point adjustment can be carried out. 

To carry out the adjustment press key 3 for 5 seconds. 

(Please note: A zero-point adjustment can only be carried out if  the difference between 

the value on display and the value calibrated on site is less than 2%! E.g. for the 

measuring range of –10.0 ..+350.0mbar, =>zero-point adjustment up to 7.0mbar possible) 

To recall the manufacturer`s calibration press button 3 for approx. 7 seconds.  

Note: If a zero-point adjustment was carried out, this will be signalled by the short 

displaying of „NuLL Corr“ when switching on the device. 

Configuration 

To change device settings, press 

Menu 

(key 4) for 2 seconds. This will call the configuration menu.  

Pressing key 

Menu

 jumps between the parameters. 

The parameters can be changed with   (key 2) or   (key 5).  

Quit

 (key 6) finishes the configuration and returns to standard measuring operation. 

Unit: Coice Of The Display Unit 

Choose the desired display unit, the refering unit is displayed by means of a functional arrow in 

the display. The selection is permanently stored in the device, therefore after power on the unit 

will instantly reapear. The choice depends on the used sensor. 

The unit [m] =mH

2

O is just supported by devices with [m] printed below the display! 

P.oFF: Auto Power Off Time 

The device will be automatically switched off if no key is pressed/no interface communication 

takes place for the time of the power off time.  

The power off time can be set to values between 1 and 120 min. It can be completely 

deactivated by setting the paramter to ‚P.oFF = oFF“.  

Adr: Base Address of Interface 

Up to 10 devices of the MH3xxx- handheld-family can be connected to a serial interface at 

once (depending on interface converter, e.g. GRS3105: 5 devices). To get access to each 

device the base addresses of the devices have to be different. For example choose 01 for the 

first, 11 for the second device and so on.  

MPa

Pa

bar

mbar

kPa

Tara

AL Logg

SL

PSI

mbar

MPa

Pa

bar

mbar

kPa

Tara

AL Logg

SL

PSI

MPa

Pa

bar

mbar

kPa

Tara

AL Logg

SL

PSI

mbar

Содержание MH3160-07

Страница 1: ...H Co KG Struthweg 7 9 34260 Kaufungen Germany 1 8 Tel 49 5605 803 0 Fax 49 5605 803 54 email info SIKA net web www SIKA net User s Manual Handheld Digital Precision Manometer MH3160 07 10 0 to 350 0 m...

Страница 2: ...vice and sensors have to be handled with care don t throw hit etc Protect plugs and sockets from soiling 3 If the device is transported from a cold to a warm environment condensation may cause in a fa...

Страница 3: ...ke sure that the operating voltage stated at the power supply is identical to the mains voltage Connections Connection for pressure tubes higher pressure lower pressure Interface Connection for el iso...

Страница 4: ...e parameters can be changed with key 2 or key 5 Quit key 6 finishes the configuration and returns to standard measuring operation Unit Coice Of The Display Unit Choose the desired display unit the ref...

Страница 5: ...e it when necessary System error Disconnect battery and power supplies wait shortly then reconnect No display or confused characters device does not react on keypress Device defective Return to manufa...

Страница 6: ...nufacturer is capable to do efficient recalibration if necessary to get results of highest accuracy The Serial Interface By means of the serial interface and a suitable electrically isolated interface...

Страница 7: ...6 Read min memory 200 Read min display range 7 Read max memory 201 Read max display range 12 Read ID number 202 Read display range unit 174 Clear min memory 204 Read display range decimal point 175 C...

Страница 8: ...ce for the time of the power off delay The power off delay can be set to values between 1 and 120 min it can be completely deactivated Display 2 four digit LCDs 12 4mm high and 7 mm high for measuring...

Отзывы: