Safety instructions
— Series VMI
- 24 -
© SIKA • Ba_VMI • 12/2020
1 Safety instructions
Before you install the device, read through this operating manual carefully. If the instructions con-
tained within it are not followed, in particular the safety guidelines, this could result in danger for
people, the environment, and the device and the system it is connected to.
The device corresponds to the state-of-the-art technology. This concerns the accuracy, the
operating mode and the safe operation of the device.
In order to guarantee that the device operates safely, the operator must act competently and
be conscious of safety issues.
SIKA provides support for the use of its products either personally or via relevant literature.
The customer verifies that our product is fit for purpose based on our technical information.
The customer performs customer- and application-specific tests to ensure that the product is
suitable for the intended use. With this verification all hazards and risks are transferred to our
customers; our warranty is not valid.
Intended use
The magnetic inductive flow sensor VMI must only be used for measuring and metering liq-
uids with a minimum conductivity of 50 μS/cm.
WARNING
The magnetic inductive flow sensors of the VMI series are no safety components
in accordance with Directive 2006/42/EC (Machine Directive).
Never use the device as a safety component.
The operational safety of the device supplied is only guaranteed by intended use. The speci-
fied limits (
§ 9 "Technical data") may under no circumstances be exceeded.
Before installation, check whether the wetted materials of the device are suitable for the liquid
used (
MEASURING TUBE EMPTY (OR PARTIALLY FILLED) /
CONDUCTIVITY TOO LOW
The green LED may blink irregularly if the measuring tube of the VMI is empty or
partially filled or if the conductivity of the fluid being used is too low. Random
pulses will be present at the output, but they do not represent an actual flow.
Ensure that the measuring tube of the VMI is always completely filled
(
§ 4.1 "Installation instructions").
Ensure that the co
nductivity of the fluid is at least 50 μS/cm.