Instruction  Manual  VisGuard 2 E xtractive 

Commissioning 

26 

14265E/1 

 

 

7

 

Commissioning 

 

 

Due to the wide range of different sampling system versions, it is not  possible to provide  de-
tailed guidelines and principles for commissioning. If plausible measuring values are seen and 
no  malfunction messages are shown on the control  unit, it  can be assumed that the samp-
ling system is working correctly. 

 
Proceed with the initial start-up  in accordance  with the  following table: 
 

 

WORKSTEP 

ADDITIONAL  INFO / IMAGES 

1. 

Check that  the sample transport  is correct and 
the sampling system has no leaks. 

 

2. 

On multiple sampling systems, ensure the cor-
rect allocation  of the  measuring point  to the 
outlet  channel of the control  device (correct 
connection  of the  electrical connections  and 
samples). 

 

3. 

Optional  components  installed for monitoring 
the flow rate in the sample lines (pressure mo-
nitor)  must be set on site while the  system is 
running.  Section 5.3.3 

 

4. 

In order  to  compensate for the  vacuum  in the 
measuring cell and dust  losses in the lines, a 
suction correction  factor may have to  be set on 
the control  unit.  This factor is based on experi-
ence and  can be determined  using Figure  4 or 
by measuring with a mobile  In-situ instrument. 

 

 

 

Figure 4: Determining the suction correction factor (KF) 

KF 

Suction  correction  factor 

Line length 

 

 

 

 

 

1 .2 0

1 .3 0

1 .4 0

1 .5 0

1 .6 0

3 0

8 0

1 3 0

1 8 0

2 3 0

2 8 0

3 3 0

3 8 0

4 3 0

4 8 0

L  [m ]

K F

Содержание VisGuard 2 Extractive

Страница 1: ...Document number 14265E Version 1 Valid from S N 704010 SW V529 INSTRUCTION MANUAL VisGuard 2 Extractive Sampling System Visibility Monitor Dust Monitor...

Страница 2: ...T PHOTOMETER AG subject to technical changes without notice 4 2018 SIGRIST PHOTOMETER AG Hofurlistrasse 1 CH 6373 Ennetb rgen Switzerland Tel 41 41 624 54 54 Fax 41 41 624 54 55 info photometer com ww...

Страница 3: ...er KW5 11 3 General safety points 12 3 1 Dangers when properly used 12 3 2 Residual risk 13 3 3 Warning and danger symbols on the instrument 13 3 4 Preventing undesirable online access attempts 14 4 D...

Страница 4: ...rd 2 Extractive 30 9 Troubleshooting on additional components 31 9 1 General troubleshooting 31 9 2 Malfunctions on the sample heater KW5 32 9 3 Malfunctions on the valve unit ST2MP3 32 9 4 Malfunctio...

Страница 5: ...installers who are responsible for the installation and operation of the mea suring instrument This document supplements the existing Instruction Manual for the Vis Guard 2 and may only be used in co...

Страница 6: ...may be operated only by trained technical personnel who have read and un derstood the content of the Instruction Manual 1 10 Declaration of conformity Current technological principles were followed in...

Страница 7: ...SIGRIST PHOTOMETER AG has not expressly recommended Improper changes have been performed to the components The components have been operated outside the defined specifications 1 13 Meaning of the saf...

Страница 8: ...ng of the pictograms All pictograms used in this document are explained below Additional information about the current topic Practical procedures when working with the VisGuard 2 Extractive Manipulati...

Страница 9: ...in the corresponding drawings Sampling type Line length Drawing Single 0 5 m VISGUARD2 GSS5 TU Single 5 30 m VISGUARD2 GSS30 TU Single 30 500 m VISGUARD2 GSS500 TU Multiple 30 200 m lengths on request...

Страница 10: ...E4n Ambient temperature Max 40 C Medium temperature Max 80 C Dimensions W x H x D 358 mm x 357 mm x 375 mm For more details see drawing KTNPI G1 MB Section 16 2 3 2 Technical data for the fan KZTN2 Da...

Страница 11: ...W x H x D 744 mm x 730 mm x 153 mm for six channels For more details see drawing ST2MP3 MB Section 16 2 3 4 Technical data for the sample heater KW5 Data Values Service voltage 115 or 230 VAC 50 60 H...

Страница 12: ...ual after the instrument changes hands Operating the instrument without knowledge of the Instruction Manual may lead to injuries to persons and damage to the instrument If the instrument changes hands...

Страница 13: ...rm the specified servicing duties 3 3 Warning and danger symbols on the instrument WARNING There are no warning or danger symbols on the instrument Users must ensure that they observe the safety measu...

Страница 14: ...et then any In ternet user can in principle access your instrument and change the configuration Please note the following points to prevent this Never connect the instrument directly to the Internet O...

Страница 15: ...the measuring instrument The correct dimensioning and routing of the lines are decisive for this part of the measuring system When mul tiple sampling is carried out a valve unit is installed downstre...

Страница 16: ...ly suctioned in by a valve unit with two or more channels Only the sample that is to be measured at the specific time is released by the corresponding valve and allowed to enter the measuring instrume...

Страница 17: ...r can be attached to a back pressure nozzle for monitoring the sample flow Section 5 3 4 The main flow is much greater than the sample flow through the measuring instrument Thanks to the high flow vel...

Страница 18: ...y a suction fan A suction fan with higher or lower suction power is used depending on the sampling system A filter can be optionally installed upstream of the suction fan as shown in drawing KZTN3 MB...

Страница 19: ...llation drawing KZTN8 1 MB Splash guard for surface installation drawing KZTN8 MB 5 2 Sample transport 5 2 1 General information The goal of sample transport is to feed the gas sample through the meas...

Страница 20: ...for external instruments KZTN9 can be integrated in the sampling system Using this connection some of the sample is made available to other measuring in struments in order to determine CO NOX etc The...

Страница 21: ...dividually 5 3 3 1 General information on mounting the back pressure nozzle and pressure monitor The back pressure nozzle is inserted into the sample flow at the intended measuring point using hose co...

Страница 22: ...and wait for around 20 seconds until the pressure conditions have stabilized 4 Slowly turn the setting screw clockwise and count the number of turns until a malfunction is signaled Determine the thres...

Страница 23: ...stalled up stream to filter larger particles out of the sample flow Installation of a filter also prevents shock waves from spreading from the suction fan into the measuring cell The filter is install...

Страница 24: ...ning in the purging air independently On sampling systems version 0 5 m Section 2 1 that use suction fans with a lower power KZTN2 the purging air is tapped on the suction fan outlet via a flow divide...

Страница 25: ...switch a suitable disconnection device switch plug should be installed near the service voltage It must be designated and easily accessible The power supply must have a back up fuse with a max trippin...

Страница 26: ...systems ensure the cor rect allocation of the measuring point to the outlet channel of the control device correct connection of the electrical connections and samples 3 Optional components installed f...

Страница 27: ...ctly Every 6 mon ths Operator Replace the suction fan filter cartridge Section 8 4 Measure for suction fan main tenance providing a suction fan is installed in your system Annually Operator Clean the...

Страница 28: ...e connections are seated correctly and fully func tional 3 Noise inspection Check for whistling or gurgling noises Such noises often indicate a leak in the system 4 Check the measuring values Measurin...

Страница 29: ...the interval increases accordingly WORKSTEP ADDITIONAL INFO IMAGES 1 Remove the four knurled screws and filter co ver as shown in drawing KZTN3 MB 2 Remove the filter cartridge and replace it 3 Faste...

Страница 30: ...ng the purge air filter on the VisGuard 2 Extractive The purging air filter can be replaced as follows WORKSTEP ADDITIONAL INFO IMAGES 1 Remove the hose clamp arrow 2 Remove the old purge air filter i...

Страница 31: ...Section 7 Measuring value is too high Incorrectly set suction correc tion factor Section 7 Soiled clogged sample lines Section 8 5 No sample heating fog build up Section 9 2 Malfunction in flow rate m...

Страница 32: ...e must be present or zero if the measured measuring point is currently active If this is not the case examine the controller for the cause documentation for the control unit Interruption in the electr...

Страница 33: ...uit breaker has triggered If a motor circuit breaker is installed check whether it has triggered If it has triggered reset the circuit breaker If the circuit breaker continues to trigger the suction f...

Страница 34: ...ple heater from the sample inlet of the photometer Warning The heater can be very hot af ter operation and may cause burns 4 Insert the new sample heater 5 Attach the hose to the sample inlet 6 Reatta...

Страница 35: ...1 Number all electrical cables and hose connections 2 Interrupt the power supply to the valves 3 Make a sketch of all connections to avoid confu sion afterwards 4 Open all the valves 5 Remove the ele...

Страница 36: ...NFO IMAGES 1 Interrupt the power supply to the suction fan If a motor protecting switch is installed upstream of the suction fan also switch this off 2 Number all electrical cables and hose connec tio...

Страница 37: ...resentatives is available online at www photometer com Please have the following information ready when you contact a SIGRIST service point or customer service The serial number of the VisGuard 2 Extr...

Страница 38: ...the components There are no special requirements for storing the components However please note the following information The VisGuard 2 Extractive and associated components contain electronic parts...

Страница 39: ...le If the original packaging is no longer available note the following information Before packaging close the openings of the instrument with adhesive tape or plugs so that no packaging materials can...

Страница 40: ...OSSIBILITIES Packaging Cardboard wood paper Reuse as packaging material local disposal center incinerati on plants Protective foils polystyrene shells Reuse as packaging material recycling Electronics...

Страница 41: ...this documentation and their article numbers are listed in the follo wing table ARTICLE NUMBER NAME REMARKS 108710 Filter cartridge insert for air filter 0 5 and 5 30 m versions Section 8 4 112407 Pu...

Страница 42: ...Instruction Manual VisGuard 2 Extractive Appendix 42 14265E 1 16 Appendix...

Страница 43: ...Appendix Instruction Manual VisGuard 2 Extractive 14265E 1 43 Thi s page i si nt ent i onal l ybl ank...

Страница 44: ...20 F Fan filter cartridge replace 29 Fan KTNPI G1 10 Fan KZTN2 10 Fan technical data 10 Filter 18 Flow control 21 Flow divider 24 Flow rate monitoring 26 Fog 23 I Improper use 7 Incorrect use 7 Inner...

Страница 45: ...mpling points check 15 28 Service center 37 Servicing duties 27 Spare parts 41 Splash guard 28 ST2MP3 20 ST2MP3 malfunctions 32 Storage 38 Storage location 6 Suction fan 18 23 Suction fan malfunctions...

Страница 46: ...SIGRIST PHOTOMETER AG Hofurlistrasse 1 CH 6373 Ennetb rgen Switzerland Tel 41 41 624 54 54 Fax 41 41 624 54 55 info photometer com www photometer com...

Отзывы: