Instruction Manual OilGuard 2 Ex
Electrical installation
36
13003E/3
5.3
Opening and closing the door on the OilGuard 2 Ex
DANGER!
Opening the photometer in potentially explosive areas.
Opening the photometer in potentially explosive areas can lead to an explosion.
The instrument may be opened only after the service voltage has been interrupted and a
period of at least ten minutes has elapsed.
The OilGuard 2 Ex has two locks on the door. A special key is required for opening and
closing the door.
Opening the OilGuard 2 Ex
Closing the OilGuard 2 Ex
Insert the special key in the locks and turn
clockwise
.
Insert the special key in the locks and turn
counter-clockwise
.