Decommissioning/storage
Instruction Manual LabScat 2
12112E/2
51
12
Decommissioning/storage
DANGER!
Pulling out or inserting plug of the power unit.
Pulling out or plugging in the power unit can be extremely dangerous. The instrument may
also be damaged. Local regulations must be observed at all times.
The aim of decommissioning is to prepare the instrument properly for storage.
PROCEDURE
ADDITIONAL INFO
1.
Interrupt service voltage to the photometer and pull out the
power unit plug.
2.
Empty water bath.
Section 4.4
3.
Remove water inlet and outlet.
Section 4.2
4.
Thoroughly clean and dry the instrument.
Section 9.2
5.
Close water bath cover.
6.
Make sure that the USB cover is mounted.
Storage:
There are no special requirements for storing the instruments. However, please note the
following information:
The instrument contains electronic components. Storage for such components must ful-
fill the usual conditions. It is important to ensure that the storage temperature is be-
tween -20 and +50
℃
.
All components that come into contact with the sample during operation have to be dry
and clean for a long time before being put into storage.
The instrument and accessory parts must be protected from weather factors, condensing
humidity, and aggressive gases.