Instruction Manual DualScat Ex
40
10118E/5
Message
Means
Possible causes
Current (1/2)
The reading output (1
or 2 in the case of the
dual-angle instrument)
is malfunctioning.
-
Open connection terminals at
the reading output
Section 3.2
-
Open current loop of the read-
ing output
Systemfault
An internal system
fault has occurred.
-
The parameter memory has lost
its data
Service technician
-
The fault history is overfilled
Service technician
Temperature
Excessive temperature
in the Ex enclosure.
-
Cooling system defective or
lacking
-
Medium temperature too high
6.3
Carrying out a sensor check
The sensor check is an internal photometer plausibility check that, with the
factory setting, is carried out automatically once a day. The checking interval
can be changed without restriction or switched off entirely. Irrespective of this
automatic mode, the sensor check can be initiated at any time either manually
or by means of external control signals.
Reference Handbook
Action
Display (Example)
Remarks
1.
Access code
> ▓00000 <
If no individual access
code was set, proceed to
step 3.
2.
Enter code:
/
change number
/
change place
Access code
> ...... <
Enter your own access
code here.
3.
* SENSOR *
* CHECK *
Instrument in service
mode.
4.
> Start check <
No
5.
Start check
> No <
Activate editing mode.
6.
Start check
> Yes <
7.
Check
running ...
8.
Wait max. 45 s
REF 90: 25:
100 100 100
Instrument runs down
successively to 010.
Sensor check
Содержание DualScat Ex
Страница 2: ... SIGRIST PHOTOMETER AG Subject to change without notice 9 2016 ...
Страница 6: ...Instruction Manual DualScat Ex iv 10118E 5 ...
Страница 53: ...Instruction Manual DualScat Ex 10118E 5 47 ...
Страница 54: ...Instruction Manual DualScat Ex 48 10118E 5 11 Appendix ...
Страница 55: ...Instruction Manual DualScat Ex 10118E 5 49 ...