Instruction Manual AquaScat S
Servicing
80
13798E/3
9.4
Cleaning the AquaScat S
The following procedure describes how to clean the measuring cell and how to check the
condition of the AquaScat S:
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Stop the sample flow and remove the AquaS-
cat S from the line according to Section 4.
It is dangerous to remove the pho-
tometer without first emptying the sam-
ple pipe:
The instrument must not be removed from
an unemptied or partly emptied pipe since
this can allow the sample medium to flood
out, causing material damage or personal
injury as a result.
If aextractable assembly is
used, the sample flow must not
be interrupted.
2.
Loosen the two screws (circles) and remove the
cover from the sensor head.
3.
Rinse out the absorber (circles).