background image

2. Specifi cations

General

Compatibility: 

Signet 5075 Totalizing Monitor

 

Signet 5500 Flow Monitor

 

Signet 5600 Batch Controller

 

Signet 8550 series Flow Transmitter

 

Signet 8900 Multi-Parameter Controller

Operating Range:
•  -1V sensor: 

100 to 2000 mL/m (0.026 to 0.528 U.S. gpm)

 

 

(This sensor available in Europe only)

•  -2V sensor: 

400 to 2800 mL/m (0.105 to 0.740 U.S. gpm)

•  -3V sensor: 

700 to 4200 mL/m (0.185 to 1.123 U.S. gpm)

•  -4V sensor: 

1300 to 6000 mL/m (0.343 to 1.585 U.S. gpm)

• -6V 

sensor: 

3200 to 12000 mL/m (0.845 to 3.170 U.S. gpm)

Accuracy: 

±2% of reading  

Repeatability: 

±0.25% of full range

Viscosity range: 

0.2 to 20.0 centistokes

Pipe connections:  G 

1

/

4

 in. ports, 

1

/

4

 in. NPT (male) pipe adapters 

(2 included)

Wetted Materials

• Housing: 

PVDF

• Flow 

insert: 

PTFE

•  Quad ring seal: 

FPM

• Rotor: 

PVDF

•  Pipe thread adapters:  PVDF
•  Suitable for clean fl uids only

Electrical

Power: 

5 to 24 VDC ±10%, regulated,10 mA max. 

Output Type: 

Open-collector, sinking, 10 mA max.

Cable Length: 

7.6m (25 ft), can be extended up to 300 m  

 

 (1000 

ft) 

Cable type: 

2-conductor shielded, twisted-pair, 22 AWG

SAFETY INSTRUCTIONS

1.  Do not remove from pressurized lines.
2.  Confirm chemical compatibility before use.
3.  Do not exceed maximum temperature/pressure specifications.
4.  Wear safety goggles or faceshield during installation/service.
5.  Do not alter product construction.
6.  Failure to follow safety instructions could result in severe personal injury.

1. Description

The Signet 2507 Mini Flow Sensor contains a free-running rotor 
which is driven by the fl uid fl ow. Within the given measurement 
range, the rotational speed of the rotor is proportional to the 
fl uid fl ow rate. Permanent magnets built into the rotor actuate an 
electronic switch in the top of the sensor generating a square-
wave output signal proportional to fl ow rate. Both opaque and 
transparent fl uids can be measured from 0.2 to 20.0 centistokes. 

Wetted sensor parts are constructed of PVDF and FPM, making 
the sensor suitable for use with most process fl uids, including 
most acids, bases, light oils, and solvents. 

Signet 2507 Mini Flow Sensor

FLOW 

Cable plug 

25 ft cable

Pipe fitting adapter 

1

/

4

 x 

1

/

4

 in. NPT 

(2 included) 

1

/

4

 in.  

port threads 

Upper body with  
electronic switch 

Lower body with  

built-in rotor 

Polar organic solvents (i.e., ketones and chlorinated 
hydrocarbons) and aromatic hydrocarbons are not 
compatible with this sensor.

Max. Temperature/Pressure Rating

5.5 bar @ -18 °C (80 psi @ 0 °F)

• 

5.5 bar @ 24 °C (80 psi @ 75 °F)

• 

3 bar @ 120 °C (45 psi @ 248 °F)

• 

Pressure Drop Across Sensor vs. Flow Rate

Shipping Weight:

  0.115 kg (0.25 lb)

Standards and Approvals

CE

• 

Manufactured under ISO 9001 and ISO 14001

• 

China RoHS (Go to www.gfsignet.com for details)

0

20

40

60

80

100

-40

0

40

80

120

160

200

240

280

°C

-40         -18            4           27            49           71           93         116        138        

°F

6.9

5.5

4.1

2.8

1.4

psi

bar

0

20

40

60

70

80

0.25

0.50

0.75

1.00

1.25

1.50

3.63

7.25

10.88

14.50

18.13

21.76

PSI

BAR

Flow rate (%) of full scale

-2V

-3V

-4V
-6V

3-2507.090 

 

Rev. P  02/10 

 

English

*3-2707.090*

English

WARNING!

Содержание 2507

Страница 1: ...personal injury 1 Description The Signet 2507 Mini Flow Sensor contains a free running rotor which is driven by the fluid flow Within the given measurement range the rotational speed of the rotor is proportional to the fluid flow rate Permanent magnets built into the rotor actuate an electronic switch in the top of the sensor generating a square wave output signal proportional to flow rate Both op...

Страница 2: ... 4 in straight threads to 1 4 in NPT pipe threads Hand tighten only Apply 1 2 turns of PTFE sealing tape to all threaded connections to prevent leaks Installation Hints Avoid vibrations and shocks Avoid solids in the fluid Install a filter or line strainer upstream to protect sensor 1 2 Sensor Ground silver wire Sensor Power 5 to 24 VDC black wire Sensor Signal Output Red wire Max cable length 300...

Страница 3: ...Assembly Tabs 3 FLOW Inlet port Outlet port flow insert Rotor Quad ring seal WARNING Do not use tools of any kind on the sensor body or port connections Hand tighten only Excessive force will damage sensor 6 Replacing The Flow Insert Sensor range can be modified by changing the flow insert The sensor must be removed from service and disassembled prior to installing the new flow insert See Specific...

Страница 4: ... Rotor 2507 3 2507 080 3 198 801 547 Quad Ring 2507 3 2507 080 5 198 801 508 DIN Connector 2507 3 2507 081 1 198 801 548 1 mm Insert available in Europe only 3 2507 081 2 198 801 502 2 mm Insert 3 2507 081 3 198 801 503 3 mm Insert 3 2507 081 4 198 801 558 4 mm Insert 5523 0222 159 000 392 Cable per foot 2 cond w shield 22 AWG Ordering Information Sensor Part Number 3 2507 100 Mini flow low flow s...

Отзывы: