Signametrics
8
2.0 Specifications and Feature Table
Function SM4020
Standard
Scanner
SM4022
Instrumentation
Scanner
SM4040
Standard
Scanner
SM4042
Instrumentation
Scanner
Number of differential channels
20
40
20
40
Number of 10:1 groups
two
four
two
four
Scanning Arrangement
Two groups of 10:1 differential
Four groups of 10:1 differential
Thermal EMF offset (
µ
V)
25 1.5 25 1.5
Maximum Switching DC Voltage (V)
110
220
110
220
Maximum Switching AC Voltage (V)
110
250
110
250
Maximum Switching Current (A)
1
1
1
1
Maximum Switching Current (A)
1
1
1
1
Typical inter-channel Capacitance
15pf
15pf
15pf
15pf
Insulation between open contacts (
M
Ω
,)
>100 >1,000 >100 >1,000
Insulation; contacts to coils
(M
Ω
)
>100 >1,000 >100 >1,000
Insulation; adjacent channels
(M
Ω
)
>100 >1,000 >100 >1,000
No Load Life
2X10
7
10
8
2X10
7
10
8
Loaded Life @ 50Vdc, 0.1A
3X10
5
10
6
3X10
5
10
6
Trigger input
√
√
√
√
Trigger output
√
√
√
√
Thermocouple Cold Junction Capable
√
√
Typical Closure Time (ms)
12
4
12
4
Typical Release Time (ms)
5
2
5
2
Actuation Time (ms) [1]
15
5
15
5
Actuation Time Settable Range (ms)
1 to 850 in 0.25 steps
AutoScan Period Range (ms)
1 to 850 in 0.25 steps
Available Configurations
2-wire, 4-wire,
universal, two-
groups
2-wire, 4-wire, 6-
wire, universal,
two-groups, four-
groups
2-wire, 4-wire,
universal, two-
groups
2-wire, 4-wire, 6-
wire, universal,
two-groups, four-
groups
Available Scan Groups
A and B
A, B, C and D
A and B
A, B, C and D
Isolated Relay Coil Drive
√
√
Maximum number of relays closed
All
All
All
40
High Ohms range 1,000 Meg
√
[1] Actuation time is the time it takes to open any closed relays, and close the selected relay.
Table 2-1. Specifications and Feature table.
2.1 Trigger Input
Input Characteristics
•
Input
LED (nominal 1.7V drop at 1mA) with a series 1k
Ω
resistor.
•
Input Signal requirements
>2.5 V, < 10 V to activate. < 1V to deactivate.
•
Isolation
Optically Isolated from all other circuitry. Common line with Trigger Output.