SIGMATEK C-DIAS Series Скачать руководство пользователя страница 1

 

C-DIAS TEMPERATURE SENSING MODULE 

CAI 888 

19.04.2012 

Page 1 

 

C-DIAS Temperature Sensing Module 

CAI 888

 

8 x Thermocouple inputs  
8 x digital 24 V 

 
 
This  analog  input  module  is  used  for  the  recording  of 
temperatures. 
 
The following thermocouple inputs can be connected: 
 
Fe-CuNi (Type J) ... 0 – 600 °C 
NiCr-Ni (Type K) ... 0 – 800 °C 
FeCu-Ni (Type L) … 0 – 600 °C 
 
In  addition,  8  digital  +24  V  outputs  (positive  switching)  are 
available. The supply voltage is monitored for under voltage. 
 
 

Technical Data 

 
Analog channel specifications 
 

Number of channels 

Measurement range 

Type J (0 - 600 

°

C; 0 – 33.1 mV) 

Type K (0 - 800 

°

C; 0 – 33.3 mV) 

Type L (0 - 600 

°

C; 0 – 33.7 mV) 

Measurement value 

0 – 6000 or  

0 – 8000 

An open input returns a value of -

2147483631 

Resolution 

0.15 °C / 12 Bits 

Conversion time per channel 

1 ms 

Applicable sensor type 
according to 

DIN 43710 

FeCu-Ni (Type J) 

NiCr-Ni (Typ K) 

FeCu-Ni (Typ L) 

Input filter 

0.5 s 

Low pass 3 system 

Common mode range 

±10 V 

Sensor type for cold junction 

KTY10-62 

Cold junction 

-20 to 80 °C 

Precision of Analog channel 
measurement 

±0.70 % of the maximum measurement value 

(with use of KTY10-62 terminal compensation sensors) 

Содержание C-DIAS Series

Страница 1: ...voltage Technical Data Analog channel specifications Number of channels 8 Measurement range Type J 0 600 C 0 33 1 mV Type K 0 800 C 0 33 3 mV Type L 0 600 C 0 33 7 mV Measurement value 0 6000 or 0 80...

Страница 2: ...C DIAS TEMPERATURE SENSING MODULE Page 2 19 04 2012 Frequency Curve Measurement error due to an internal sensor resistance error Measurement error due to an internal sensor resistance Resistance Ohm E...

Страница 3: ...lay 100 s Turn off delay 150 s Max Braking energy per output 1 J Electrical requirements Supply voltage of the digital outputs X6 18 30 V DC Current consumption of the digital output supply X6 Corresp...

Страница 4: ...ditions Storage temperature 20 85 C Operating temperature 0 60 C Humidity 0 95 uncondensed EMV stability According to EN 61000 6 2 industrial area EMV noise emission According to EN 61000 6 4 industri...

Страница 5: ...C DIAS TEMPERATURE SENSING MODULE CAI 888 19 04 2012 Page 5 Mechanical Dimensions 104 10 dimensioning incl covers 129 109 2 24 90...

Страница 6: ...CAI 888 C DIAS TEMPERATURE SENSING MODULE Page 6 19 04 2012 Connector Layout...

Страница 7: ...ture input 1 3 Temperature input 2 4 Temperature input 2 5 Temperature input 3 6 Temperature input 3 Pin Function 1 Temperature input 4 2 Temperature input 4 3 Temperature input 5 4 Temperature input...

Страница 8: ...2012 X4 Digital outputs plug 1 4 Pin 1 X5 Digital outputs plug 5 8 Pin 1 X6 Voltage supply for the digital outputs Pin Function 1 Output 1 2 Output 2 3 Output 3 4 Output 4 Pin Function 1 Output 5 2 Ou...

Страница 9: ...cable connectors X1 X3 6 pin Weidm ller plug B2L3 5 6 AU X4 5 Phoenix Contact FK MCP 1 5 4 ST 3 5 X6 Phoenix Contact FK MCP 1 5 2 ST 3 5 The complete C DIAS CKL 043 connector set with spring terminals...

Страница 10: ...DULE Page 10 19 04 2012 Status Displays LED number LED color Definition 1 Yellow Output 1 2 Yellow Output 2 3 Yellow Output 3 4 Yellow Output 4 5 Yellow Output 5 6 Yellow Output 6 7 Yellow Output 7 8...

Страница 11: ...ls alloys this is called the Seebeck effect This voltage exists therefore not only at the measurement point where it is desired but also at the coupling between the Thermocouple conductors and the cop...

Страница 12: ...ermocouple 0V rail Temp comp Temp comp 1 2 5 6 5 6 1 2 X3 If the sensor is galvanically separated GND must be bridged with the Temp input at the terminal With non separated sensors a sufficient mass c...

Страница 13: ...comp 1 2 5 6 KTY 10 62 Intermediate terminal Thermocouple 5 6 1 2 0V rail If the sensor is galvanically separated GND must be bridged with the Temp input at the terminal With non separated sensors a...

Страница 14: ...sufficient for the maximum total current The outputs and can be turned off by turning off the 24 V supply voltage Applying power to an output whose supply voltage exceeds 0 7 V is not allowed Inductiv...

Отзывы: