Laboratory centrifuge Sigma 2-7, Sigma 2-7 Cyto
Laboratory centrifuge Sigma 2-7 IVD
6 Using the centrifuge
Version 07/2015, Rev. 1.12 of 16/07/2019
• sb
29 / 68
Translation of the original operating manual, part no. 0702802
NOTE
Pay attention to the marking of the centrifuge (see the illustration below)!
Safety indications on the centrifuge must be kept readable at all times. If
necessary, they must be replaced.
Fig. 7: Safety indication on the centrifuge: Loading of a swing-out rotor
6.2.2.3
Adapters
In order to ensure easy handling, even if vessels of various sizes are used,
carrier systems were developed.
• Load the opposite adapters with the same number of vessels and with
the same weights in order to avoid imbalance.
• If all of the compartments of a carrier are not used, the buckets must be
loaded evenly. Loading the edges of a bucket only is not permissible.
6.2.2.4
Vessels
General
• Load the vessels outside of the centrifuge. Liquids in the bores of the
rotor cause corrosion.
• Fill the vessels carefully and arrange them according to their weight.
Imbalances result in the excessive wear of the bearings.
• Follow the safety instructions and hazard warnings (see chapter 3 -
Glass vessels
• When using glass vessels, the maximum value of 4,000 x g must not be
exceeded (except special high-strength glass vessels; please refer to
the information provided by the manufacturer).
• When inserting 100 ml glass tubes (part no. 15100) into the buckets, a
rubber cushion (part no. 16051) is mandatory.
CAUTION
If the glass tubes (part no. 15100) are used without a rubber cushion, they
may break!